7
Barnstead |Thermolyne • 800-553-0039
Mounting and Utility
WARNING
Do not mount E-pure
®
directly over equipment that requires electrical service. Routine
maintenance of this unit may involve water spillage and subsequent electrical shock
hazard if improperly located.
WARNING
Ensure that the equipment is connected to electrical service according to local and
national electrical codes. Failure to properly connect may create a fire or shock hazard.
WARNING
This device is to be used with water feeds only. Sanitizing/cleaning agents must be used
in compliance with instructions in this manual. Failure to comply with the above could
result in explosion and personal injury.
CAUTION
Do not connect unit to electrical service until instructed to do so.
CAUTION
Wall composition, condition and construction, as well as fastener type must be
considered when mounting this unit. The mounting surface and fasteners selected must
be capable of supporting a minimum of 145 pounds. Inadequate support and/or fastener
may result in damage to mounting surface and/or equipment. If you are unsure of
mounting surface composition, condition and construction or correct fasteners, consult
your building maintenance group or contractor.
Screws and fasteners required for wall mounting are not supplied with the unit. The recommended
dimensions for mounting hardware to support your E-pure
®
are:
Body diameter - 1/4" maximum
Head diameter - 3/8" minimum
Length - 1” typical
The E-pure
®
comes completely assembled. The only requirements are to secure the unit to the wall and
have a source of feedwater and electrical service nearby.
The E-pure
®
comes complete with a 6 (six) foot power cord that must be plugged into a suitable outlet.
(See specification plate for electrical requirements).