24
ProG LiSt (Program Edit List)
PrG.n: Press the UP or DOWN ARROW to select the
program number (program number will be displayed in
lower display on 4x16 programmable models only.)
Hb: Press the UP or DOWN ARROW to select the
holdback type [OFF (disables holdback), Lo (deviation
low holdback), Hi (deviation high holdback) or bAnd
(deviation band holdback)] for the entire program. The
controller will default to “OFF.”
Hb.U: Press the UP or DOWN ARROW to select the
holdback value (in display units).
rmP.U: Press the UP or DOWN ARROW to toggle
between ramp units (SEc, min or Hour). Controller will
default to “SEc.”
dwL.U: Press the UP or DOWN ARROW to toggle
between dwell units (SEc, min or Hour). Controller will
default to “SEc.”
Cyc.n: Press the UP or DOWN ARROW to set the
number of program cycles (1 to 999 or cont). The con-
troller will default to “cont.”
SEG.n: Press the UP or DOWN ARROW to select the
segment number (1-8 in 8 segment models, 1-16 in
4x16 models).
tYPE: Press the UP or DOWN ARROW to select the
segment type [End (end of program), rmP.r = ramp rate
(ramp to a specified setpoint at a set rate), rmp.t =
ramp time (ramp to a specified temperature in a set
time), dwEll (to maintain a constant temperature for a
set time), StEP (climb instantaneously from current to
specified temperature), cALL (to call a program as a
subroutine, available only on 4x16 programmable mod-
els)]. The controller will default to “End.” Other parame-
ters used with tYPE include; tGt target setpoint), Rate
(rate of temperature increase) and dur (time to target
setpoint or time to dwell).
End.t: End segment type: dwELL (dwell continuous),
rSEt (reset) and S OP (End Segment Output power
level.
8 S
EGMENT
& 4
X
16 S
EGMENT
P
ROGRAMMABLE
M
ODELS W
/OTP