1.
Turn SET-TEMP LIMIT’s power on.
The red “ALARM” light will be on.
2.
Set the limit temperature by adjusting
the dial. If necessary, use a flat blade
screwdriver to turn the dial.
Note that the temperature scale is in degrees
C and degrees F.
3.
Press the reset key to energize SET-
TEMP LIMIT’s receptacle.
Note that each time SET-TEMP LIMIT looses
power or is turned off, the “RESET” key must
be pressed to resume heating. Similarly, if the
thermocouple disconnects or open-circuits,
the thermocouple problem must be corrected.
After correcting the problem, press the
“RESET” key to resume heating.
7
Operation