33
Error Codes
ERROR0
Instrument malfunction. Call
for service
ERROR1
Blank solution exceeds
range.
ERROR2
Absorbance of tested sample
lower than that of blank
solution.
ERROR3
Absorbance of tested sample
too high.
ERROR4
Blank solution exceeds
range.
ERROR5
Instrument malfunction. Call
for service.
ERROR6
Instrument malfunction. Call
for service.
Filter Motor Error
Filter Motor is not function
ing.
Wavel. Motor Error Wavelength Motor is not
functioning.
Note
1.
Be sure to check the Blank
solution or sample solution
and WAVELENGTH first when-
ever an error message shows
on the display.
2.
After checking or correction,
press 100%T/OA button to
blank again.
Note
1.
Turn the power off and turn the
power on to re-start the sys-
tem.
2.
If the display continues to
show an error message, call
for service.