27
To Replace Type Platinel II Thermocou-
ple (Single Set Point Control Models)
a.
Disconnect furnace from power supply.
b.
Remove knob that holds furnace chamber to the
bracket and remove the furnace chamber.
c.
Remove terminal plate on back of furnace chamber.
d.
Remove two screws that secure the old thermocou-
ple and remove thermocouple by pulling straight
back.
e.
Insert the new thermocouple until tip extends ap-
proximately 1/4" into heating chamber.
f.
Connect the blue and yellow beaded thermocouple
lead to the terminal across from yellow thermocou-
ple extension wire and the other thermocouple lead
to the remaining terminal. (See figure 2)
g. Insert the screw on the terminal plate through the
mounting bracket and secure with knob.
h. Reconnect furnace to power supply.
M
AINTENANCE AND
S
ERVICING