56-101-00930 Revision H
Page 13
3. Connect the
standard leads to the Test Set’s TEST connector.
4. Press the Test Set PWR pushbutton.
5. Set the FUNCTION switch to RES MEAS.
6. Set the RANGE switch to 200
Ω.
7. Short the test lead clips together to ensure a good electrical connection.
8. Press the PTS pushbutton while adjusting the SYS RES control knob for displayed system
resistance to be simulated.
9. Release PTS pushbutton.
10. Set FUNCTION switch to IND TEST.
11. Press the mV, °F or °C pushbuttons to select units. Press the selected key again for mode
(mV) or resolution (°F or °C).
12. Connect test lead clips to indicator terminals. Observe polarity: Alumel is negative (-) and
connects to the TT1200A BLACK clip. Chromel is positive (+) and connects to the TT1200A
RED clip.
13. Set the TEMP ADJ control knob for the desired test temperature as displayed on the
TT1200A display indicator. It may be necessary to use the FINE control to achieve the exact
settings.
14. Compare the readings of indicator under test with the TT1200A readings.
Note:
Indicator must agree with TT1200A reading to within manufacturers'
specifications.
15. Press the PWR pushbutton to power down the Test Set.
16. Disconnect the Test Set from the aircraft wiring.
17. Return the aircraft to its original configuration.
E.
Temperature Measurement Test Procedure
1. Access the aircraft system temperature indicator.
2. Disconnect thermocouple leads from indicator.
3.
Connect standard leads to Test Set’s TEST connector.
4. Press the Test Set PWR pushbutton.
5. Set the FUNCTION switch to TEMP MEAS.
6.
Set the RANGE switch to 20 KΩ - 0 Ω SYS RES.
7. Press the mV, °F or °C key to select units of measurements.
CAUTION
: T
HE AIRCRAFT INDICATORS OF THE D’ARSON-TYPE MUST NOT BE
CONNECTED DURING THIS TEST. D’ARSONVAL INDICATORS MAY
GENERALLY BE RECOGNIZED BY THE ABSENCE OF A MULTI PIN
CONNECTOR AND USUALLY HAVE TWO TERMINAL POSTS OR TERMINAL
SCREWS.
8. Connect test lead clips to indicator terminals. Observe polarity: Alumel is negative (-) and
connects to the TT1200A BLACK clip. Chromel is positive (+) and connects to the TT1200A
RED clip.