2.5. Cue™ Needle Tracking System Cautions (if enabled)
Caution:
Keep the needle and any ferromagnetic objects, e.g., wired undergarments, metal instruments, watches,
jewelry, Cue ™ Activator, electronics, metal bed rails, etc. at least 3 feet away from probe when
calibrating. Failure to do so may impact the accuracy of needle tracking.
Caution:
Equipment operating in close proximity may emit strong electromagnetic or radio frequency interference
which could affect the performance of this device. Avoid operating the device near pumps, cauterizers,
diathermy equipment, cellular phones, or other portable and mobile radio frequency communications
equipment. Maintain equipment separation of at least 3 feet.
Caution:
Excessive movement of the probe may result in a temporary loss of needle tracking.
Caution:
The Activator produces a magnetic field. It should be kept away from any device that is sensitive to
magnetic fields.
Caution:
Follow all care, handling and/or installation instructions included in this guide and on the packaging
provided with the Activator and needles.
Caution:
Single-use items (e.g., needle trays, needle guides, needles and sheaths/ covers) must not be reused
under any circumstances.
Caution:
Movement of the probe during calibration may result in calibration error.
Caution:
If needle tracking is not functioning properly, discontinue use of Cue™ Needle Tracking System.
Caution:
Improper technique and environmental conditions may introduce variation in accuracy.
2.6. Pinpoint™ GT Needle Technology Cautions (if enabled)
Caution:
Keep the needle and any ferromagnetic objects, e.g., wired undergarments, metal instruments, watches,
jewelry, electronics, metal bed rails, etc. at least 2 feet away from probe when calibrating. Failure to do so
may impact needle tracking.
Caution:
Equipment operating in close proximity may emit strong electromagnetic or radio frequency interference
which could affect the performance of this device. Avoid operating the device near pumps, cauterizers,
diathermy equipment, cellular phones, or other portable and mobile radio frequency communications
equipment. Maintain equipment separation of at least 5 feet (1.5m).
Caution:
Excessive movement of the probe after calibration may result in reduced needle tracking accuracy.
Caution:
During insertion, use a light touch to prevent needle deflection and/or bending. Needle deflection and/or
needle bending may reduce needle tracking accuracy.
Caution:
To ensure accuracy, do not apply force or pressure that may leverage the needle. This may result in
needle deflection, bending or altering needle trajectory.
Caution:
Movement of the probe during calibration may result in calibration error.
Caution:
If needle tracking is not functioning properly, discontinue use of Pinpoint™ GT Needle Technology.
Caution:
To avoid reduced needle tracking accuracy, prior to insertion do not modify the straightness of the needle
by bending.
Caution:
Improper technique and environmental conditions may introduce variation in accuracy.