procedure to ensure complete removal of radioactive materials. The shielding glass 
contains lead, and must be disposed of according to local regulations. 
 

Caution:  Storage requirements are temperature dependent  

 
Extra care should be taken to avoid exposing the Q

UICK

L

INK

®

 cartridges containing 

bioabsorbable S

OURCE

L

INK

 

components to excessive heat or moisture. Store 

unopened packages at room temperature. Discard open, unused cartridges, 
S

OURCE

L

INK

  

Connectors, and S

OURCE

L

INK

 

Spacers.  

 
Do not store bioabsorbable components at temperatures that exceed 40°C.  
 

Caution:  Elevated temperatures following loader steam sterilization  

 

After sterilization, allow the Q

UICK

L

INK

®

 Delivery System loader components to cool to 

room temperature prior to use.

 

 

Caution:  Avoid damaging the system or using damaged materials 

 

  Never use visibly damaged, bent or broken Q

UICK

L

INK

®

 Delivery System loaders, 

cartridges or components. This may cause system damage or jamming. 

 

  Do not force cartridges into the carriage slots. The slots are keyed to accept the 

appropriate cartridges, and minimal effort is required to properly seat the cartridges. 

 

  When handling this or any other synthetic absorbable material, care should be taken 

to prevent any damage to the material. Avoid crushing or crimping damage to the cup 
ends caused by the application of forceps or tweezers.     

 

  Ensure the needle adapter and stylet are finger-tight. Do not make any other 

adjustments to the Q

UICK

L

INK

®

 Delivery System.   

 

  Do not tamper with the mechanisms that reside beneath the carriage when removing 

the carriage from the loader. These components are aligned to rigid specifications, 
and must be handled carefully to maintain proper loader functionality. 

 

  In the event the Q

UICK

L

INK

®

 Loader or cartridges become inoperable due to damage 

or malfunction, any or all components may be removed from the cartridges and 
implanted manually. 

 

  Do not adjust the indicator in the cartridge when the cartridge is loaded. The indicator 

provides a spring bias to the components to allow for dispensing in the loader, and lifting 
the indicator and allowing it to snap closed can damage the components and cartridge. 

 

  Do not re-sterilize S

OURCE

L

INK

 

Connectors or S

OURCE

L

INK

 

Spacers either loose or 

in cartridges.  The components are supplied sterile and indicated for single use.  If 
desired, radioactive seeds may be removed from a cartridge and re-sterilized.  Do not 
re-sterilize Q

UICK

L

INK

®

 cartridges.   

 

Caution:  Limitations on use 

 

The Q

UICK

L

INK

®

 Delivery System is prescription use only.  Users should be familiar with 

procedures and techniques involving absorbable spacers and seed train products before 
using the S

OURCE

L

INK

 

Connectors.  S

OURCE

L

INK

 

Connectors are designed to provide 

accurate spacing in increments of 0.5cm with typical 4.5mm brachytherapy sources.  
Use of S

OURCE

L

INK

 

Connectors with brachytherapy sources of other lengths will result 

in atypical spacing. 

 

Caution:  Key points for use of the Q

UICK

L

INK

®

 Delivery System 

 

  Ensure that the compression handle is fully seated to the compression slide before 

assembling a seed train.  

 

  The gate button must remain depressed until the seed train is completely dispensed 

from the loader and the stylet is retracted. 

 

  Ensure the lead glass door is closed prior to assembling or dispensing seed trains. 

 

DIRECTIONS FOR USE 

 

1.  Remove the protective tab from the bottom of the Q

UICK

L

INK

®

 cartridge prior to 

use by grasping the cartridge by the cartridge body and pulling the tab out of the 
side of the cartridge.  To prevent inadvertent dispensing of a component, do not 
hold the cartridge body by the green plunger when removing the tab. 

 

2.  Insert the cartridges containing seeds, S

OURCE

L

INK

 

Connectors, and 

S

OURCE

L

INK

 

Spacers into the carriage. Ensure that the cartridges are fully 

seated in the carriage receptacle.  The carriage and cartridges are marked as 
follows to indicate the correct placement of cartridges as well as the current 
selection: 

 

 

Radioactive

 i

mplant seed Q

UICK

L

INK

®

 Cartridge 

 

5.5mm Standard S

OURCE

L

INK

 

Cartridge

 

 

5.0mm Extension S

OURCE

L

INK

 

Cartridge

 

  

0.5mm Seed-to-Seed S

OURCE

L

INK

 

Cartridge

 

 

S

OURCE

L

INK

 

Spacer Cartridge 

 

3.  Retract the compression slide handle to the right until it contacts the rear plate to 

seat the slide compression mechanism (if unseated). 

 

4.  Attach the implant needle to the Q

UICK

L

INK

®

 Delivery System by pushing the 

needle hub firmly onto the needle adapter. 

 

5.  Move the carriage to select which item to dispense by aligning the indicator with 

the symbol indicating the desired item. 

 

Note: 

The position of the Extension S

OURCE

L

INK

 

Connector in the cartridge is 

accurately represented by the icon. Therefore, when spacing at either end 

of a seed train is desired, an Extension S

OURCE

L

INK

 

Connector may 

only be used at the leading end of the train. Spacing at the trailing end of 
the train must be provided by one or more spacers. 

 

6.  Push the dispense button to dispense and move the selected item into the 

assembly track.  

 

Note: 

During normal operation, the dispense button will return to the “up” position 
after a component is dispensed. If a cartridge is empty, the button will come 
to a stop before it reaches the bottom of its range of motion.  Further 
pressing will cause the button to release from its internal mechanism to 
prevent damage to the cartridge. The button may be lifted to the “up” 
position to reseat it to the internal mechanism, and the empty cartridge may 
be removed and replaced. 

 

Note: 

The dispense mechanism is designed to completely dispense a single 
component once depressed. The button must be pressed through its 
complete range of motion – if the button is only pressed partially down, it 
will remain in the midrange position and prevent carriage motion or further 
dispensing. Complete the compression of the button to dispense the 
selected component and release the carriage for further dispensing. 

 

7.  Repeat items 4-5 until all items necessary to build the predetermined seed train 

have been dispensed. 

 

8.  Visually inspect the dispensed components, if desired, through the lead glass door 

prior to assembly of the seed train. Incorrect components can be removed as 
necessary by opening the lead glass door to access the assembly track. 

 

Note: 

If desired, the carriage mechanism may be removed from the loader to 
inspect the area beneath the carriage. To remove the carriage, depress the 
carriage release button while lifting the carriage upward. To replace, align 
the two latch prongs on the underside of the carriage with the 
corresponding receptacles in the loader and snap the carriage into place. 

 

9.  Move the carriage to the arrow symbol indicating that the Q

UICK

L

INK

®

 Delivery 

System is ready to either compress or dispense a seed train. 

 

10.  Move the handle on the releasable slide left towards the carriage to compress the 

seed train. Continue motion in this direction until the slide handle releases and 
comes to a full stop. 

 

11. Retract the slide to the right until it contacts the rear plate to seat the slide 

compression mechanism. 

 

12. The assembled seed train may be inspected through the lead glass door, and 

proper seed spacing confirmed using the assembly base ruler. 

 

13. Move the handle on the releasable slide left again towards the carriage while 

holding the gate button down in order to dispense the seed train through the 
needle adapter. 

 

14.  Retract the handle of the releasable slide fully to the right. 

 

15.  Remove the loaded implant needle by pulling the needle hub away from the needle 

adapter. 

 

TROUBLESHOOTING 

 

Condition: S

OURCE

L

INK

Connectors not connected after compression 

  Ensure compression slide has been fully reset before compression; 

  Stylet bent – replace; 

  The seed train is incorrectly built consisting of either two female or two male 

ends placed adjacent to each other;  

 Ensure the S

OURCE

L

INK

 

Connector material is not damaged preventing 

compression. 

Condition: S

OURCE

L

INK

Connectors are ejected from the loader during 

compression 

  Ensure the lead glass cover is securely closed; 

  Do not press the gate button during compression, allowing ejection of unlinked 

components through the needle adapter. 

Condition: Pushing the dispense button does not eject any items or 
produces a partial dispense 

  Ensure that the handle is not being pulled while the dispense button is being 

depressed.  Pulling on the handle during dispensing can cause the cartridge 
to become misaligned with the dispense track, resulting in a lock out or a jam; 

  Ensure the selected cartridge is fully seated in the corresponding carriage 

receptacle; 

  Remove cartridge from carriage and reinsert back into the carriage; 

  Ensure the dispense blade is not jammed - remove the carriage and inspect 

dispense blade, verify there are no loose components; 

  The selected cartridge is empty - replace with a full cartridge; 

  The dispense button is not fully depressed - press button fully; 

  If cartridge does not dispense following troubleshooting, replace cartridge. 

Condition: The assembled seed train cannot be loaded into the attached 
implant needle 

  Ensure that you hold the gate button down while sliding the releasable slide. 

  Ensure the seed train has been properly built and does not include damaged 

components 

Condition: Implant needle does not stay attached to the needle adapter 

  Ensure needle adapter is appropriate for the implant needle used. 

Condition: Completed seed train does not load smoothly into the attached 
implant needle. 

  Ensure needle adapter is appropriate for the implant needle used. 