background image

I n d i c a t i o n s

The

 PowerPICC* SV 

catheter is indicated for short or long-term peripheral access to the central venous system for intravenous 

therapy, blood sampling, power injection of contrast media, and allows for central venous pressure monitoring. For central venous 
pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used.

Catheter Size

Maximum Flow Rate

3 F Single Lumen

1 mL/sec

4 F Dual Lumen

2.5 mL/sec

C o n t r a i n d i c a t i o n s ,   Wa r n i n g s   a n d   P r e c a u t i o n s : 

The device is contraindicated whenever:

• 

The presence of device related infection, bacteremia, or septicemia is known or suspected.

• 

The patient’s body size is insufficient to accommodate the size of the implanted device.

• 

The patient is known or is suspected to be allergic to materials contained in the device.

• 

There has been past irradiation of prospective insertion site.

• 

There have been previous episodes of venous thrombosis or vascular surgical procedures at the prospective placement site.

• 

There are local tissue factors that may prevent proper device stabilization and/or access.

Wa r n i n g s :

  When using alcohol or alcohol containing antiseptics with polyurethane PICCs, care should be taken to avoid prolonged 

or excessive contact.  Solutions should be allowed to completely dry before applying an occlusive dressing.  Chlorhexidine 
gluconate and/or povidone iodine are the suggested antiseptics to use. 

  Alcohol should not be used to lock, soak or declot polyurethane catheters because alcohol is known to degrade polyurethane 

catheters over time with repeated and prolonged exposure.

  Acetone and polyethylene glycol containing ointments should not be used with polyurethane catheters, as these may cause 

failure of the device.

  This is not a right atrium catheter.  Avoid positioning the catheter tip in the right atrium.  Placement or migration of the 

catheter tip into the right atrium may cause cardiac arrhythmia, myocardial erosion or cardiac tamponade.  The risk of these 
complications may be more likely in neonatal patients.

  Intended for Single Use.  DO NOT REUSE.  Reuse and/or repackaging may create a risk of patient or user infection, 

compromise the structural integrity and/or essential material or design characteristics of the device, which may lead to device 
failure, and/or lead to injury, illness or death of the patient. 

  After use, this product may be a potential biohazard. Handle and discard in accordance with accepted medical practice and 

applicable local, state and federal laws and regulations.

  Failure to ensure patency of the catheter prior to power injection studies may result in catheter failure.

•  Use of lumens 

not

 marked “Power Injectable” for power injection of contrast media may cause failure of the catheter.

  Power injector machine pressure limiting feature may not prevent over-pressurization of an occluded catheter, which may lead 

to catheter failure.

  Exceeding the maximum power injection flow rate or setting the maximum power injector pressure limit above 300 psi, may 

result in catheter failure and/or catheter tip displacement.

 

PowerPICC* SV

 catheter indication for power injection of contrast media implies the catheter’s ability to withstand the 

procedure, but does not imply appropriateness of the procedure for a particular patient.  A suitably trained clinician is 
responsible for evaluating the health status of a patient as it pertains to a power injection procedure.

  If the artery is entered, withdraw the needle and apply manual pressure for several minutes.

  Place a finger over the orifice of the sheath to minimize blood loss and risk of air aspiration. The risk of air embolism is reduced 

by performing this part of the procedure with the patient performing the Valsalva maneuver until the guidewire is inserted into 
the needle.

  Do not use the catheter if there is any evidence of mechanical damage or leaking. Damage to the catheter may lead to rupture, 

fragmentation, possible embolism, and surgical removal.

  If signs of extravasation exist, discontinue injections. Begin appropriate medical intervention immediately.

  The fluid level in the catheter will drop if the catheter connector is held above the level of the patient’s heart and opened to air. 

To help prevent a drop in the fluid level (allowing air entry) while changing injection caps, hold the connector below the level of 
the patient’s heart before removing the injection cap.

  Central venous pressure monitoring should always be used in conjunction with other patient assessment metrics when 

evaluating cardiac function.

•  Failure to warm contrast media to body temperature prior to power injection may result in catheter failure.
•  Do not wipe the catheter with acetone base solutions, or ointment.  These can damage the polyurethane material if used over 

time.

P r e c a u t i o n s :

  Only medical practitioners licensed by law, trained and experienced in proper positioning of catheters in the central venous 

system using percutaneous entry (Seldinger technique) should place this catheter.

  Follow Universal Precautions when inserting and maintaining the catheter.

  Follow all contraindications, warnings, cautions, precautions and instructions for all infusates, including contrast media, as 

specified by their manufacturer. 

Summary of Contents for PowerPicc SV

Page 1: ...user should contact Bard Access Systems Inc to see if additional product information is available Revised date November 2010 Bard PowerPICC and StatLock are trademarks and or registered trademarks of...

Page 2: ...Remove the injection needleless cap from the PowerPICC SV catheter 2 Attach a 10 mL or larger syringe filled with sterile normal saline 3 Aspirate for adequate blood return and vigorously flush the c...

Page 3: ...Failure to ensure patency of the catheter prior to power injection studies may result in catheter failure Use of lumens not marked Power Injectable for power injection of contrast media may cause fail...

Page 4: ...et prior to use repositioning or withdrawal II During placement Do not allow device contact with sharp instruments Mechanical damage may occur Use only smooth edged atraumatic clamps or forceps Do not...

Page 5: ...ly the tourniquet above the intended insertion site to distend the vessel 3 Prepare the site according to institution policy using sterile technique 4 Drape the patient by placing the fenestrated drap...

Page 6: ...Re advance the stylet to the distal end of the trimmed catheter 11 Insert and Advance the Catheter 1 Insert the catheter and stylet as a unit into the microintroducer sheath 2 Advance the catheter sl...

Page 7: ...ter stabilization device with transparent dressing 3 Place anchor tape sticky side up under hub Wedge tape between hub and wings 4 Chevron anchor tape on top of transparent dressing Dual Lumen 1 Secur...

Page 8: ...to flushing and aspiration may be partially or completely occluded Do not flush against resistance If the lumen will neither flush nor aspirate and it has been determined that the catheter is occlude...

Reviews: