background image

10

A

B

C

OD 

(Fr)

OD 

at Zero 

(Fr)

Taper 

Length 

(cm)

3

5

8

4

6

8

5

7

8

X

X

X

A

B

C

A

B

C

A

B

C

9.  Complete Catheter Insertion

A.  Complete catheter advancement into the desired position.

 WARNING:

 (Pediatric) This is not a right atrium catheter. Avoid positioning the cath-

eter tip in the right atrium. Placement or migration of the catheter tip into the right 

atrium may cause cardiac arrhythmia, myocardial erosion 

or cardiac tamponade. The risk of these complications may 

be more likely in neonatal patients.

           

NOTE:

 Midlines should be positioned with the catheter tip 

distal to the shoulder and deltoid muscle.

           

Caution:

 The midline catheter tip location should be at or 

near the axillary line [INS, 2016].

10. Retract and Remove the Introducer Sheath

A.  Stabilize the catheter position by applying pressure to the vein distal to the  

 

  introducer sheath.

B.  Withdraw the introducer sheath from the vein and away from the site.

C.  Split the introducer sheath and peel it away from the catheter.

           Caution:

 Do not pull apart the portion of the sheath that remains in the vessel. 

To avoid vessel damage, pull back the sheath as far as possible  

 

           

and tear the sheath only a few centimeters at a time.

11. 

Remove the Stylet / T-Lock Assembly

A.  Disconnect the T-Lock from the catheter luer connector.

B.  Stabilize the catheter position by applying light pressure  

  to the vein distal to the insertion site.

C.  Slowly remove the T-Lock and stylet.

Caution:

 Never use force to remove the stylet. Resistance 

can damage the catheter. If resistance or bunching of 

the catheter is observed, stop stylet withdrawal and allow 

the catheter to return to normal shape. Withdraw both the catheter and stylet together 

Zero mark

Summary of Contents for Poly Midline

Page 1: ...1 Instructions For Use BARD...

Page 2: ...or past treatment will prevent proper device stabilization and or access Warnings General Warnings When using alcohol or alcohol containing antiseptics with polyurethane midlines care should be taken...

Page 3: ...dure for a particular patient A suitable trained clinician is responsible for evaluat ing the health status of a patient as it pertains to a power injection procedure Failure to warm contrast media to...

Page 4: ...the dose Do not infuse against resistance To reduce potential for blood backflow into the catheter tip always remove syringes slowly while injecting the last 0 5 mL of sterile saline Do not reinsert n...

Page 5: ...round any part of the catheter s tubing shaft or extension legs If using sutures to secure catheter USE THE SUTURE WINGS and make sure they do not occlude puncture or cut the catheter Do not use sciss...

Page 6: ...Erosion Through the Skin Catheter Embolism Catheter Occlusion Catheter Related Sepsis Exit Site Infection Exit Site Necrosis Extravasations Fibrin Sheath Formation Hematoma Heparin Induced Thrombocyto...

Page 7: ...lease tourniquet D Assess the selected vein to ensure the vessel size is adequate to accommodate the catheter being placed 2 Patient Position Catheter Measurement A Position the arm at a 90 angle B Fo...

Page 8: ...tylet assembly as one unit Caution Catheter stylet must be wetted prior to stylet repositioning or withdrawal D Retract the stylet to well behind the point the catheter is to be cut Caution Do not all...

Page 9: ...nit to prevent the needle from damaging or shearing the guidewire F Advance the small sheath and dilator together as a unit over the guidewire using a slight rotational motion If necessary a small inc...

Page 10: ...Stabilize the catheter position by applying pressure to the vein distal to the introducer sheath B Withdraw the introducer sheath from the vein and away from the site C Split the introducer sheath an...

Page 11: ...er to Instructions For Use on the proper use and removal The StatLock Stabilization Device should be monitored daily and replaced at least every seven days Caution The catheter must be secured in plac...

Page 12: ...be tween hub and wings Chevron anchor tape on top of transparent dressing Tape Strip Securement Procedure Single Lumen 1 Place 1st anchor tape over wings or bifurcation 2 Cover site and 1st anchor tap...

Page 13: ...emperature prior to power injec tion may result in catheter failure Warning Use of lumens not marked Power Injectable for power injection may result in catheter failure G Complete power injection stud...

Page 14: ...LLER THAN 10 mL TO FLUSH AND CONFIRM PATENCY Patency should be assessed with a 10 mL or larger syringe with sterile saline Upon confirmation of patency administration of medication should be given in...

Page 15: ...nd Toxic Reactions Adverse Events During Infusion Therapy Symposium University of California 1993 2 Findlay Steven R et al Hyperosmolar Triggering of Histamine Release from Human Basophils Journal of...

Page 16: ...daccess com Clinical Information 1 800 443 3385 USA Email medical services crbard com STERILIZE Manufacturer Do Not Re Use Non Pyrogenic Reorder Number Lot Number Do Not Re sterilize Length Use By Ste...

Reviews: