35
R5913350 /05
Nio Fusion 12MP
TG18. That's why QAWeb Agent acts as the master for all supported displays from the moment it is installed
and running. QAWeb Agent will take over from Embedded QA and overwrite any settings which were applied
by Embedded QA.
4.18.2 DICOM status report
About DICOM status report
Following information is available:
DICOM Compliance Status
(status since last compliance check)
•
Compliance status:
Shows if the current DICOM curve is compliant or not.
•
Maximum error:
Shows the maximum error of the current DICOM curve. This is the deviation compared to
a perfect DICOM.
•
Error threshold:
Shows the error threshold. This is the maximum error allowed before a DICOM
calibration is required.
•
Time elapsed since latest compliance check:
Shows the backlight runtime since last compliance check.
•
Display Function:
Shows the current display function.
•
Ambient light compensation:
Shows the ambient light compensation status.
•
Reading Room:
Shows the selected reading room.
•
Luminance:
Shows the measured luminance.
•
Black luminance:
Shows the measured black luminance.
DICOM Calibration Status
•
No calibration executed yet:
No other information is visible
•
Calibration executed:
When the calibration is executed, the following extra information is shown: Time
elapsed since latest calibration, Display Function, Ambient Light Compensation and Reading Room.
Current DICOM Settings
•
Display Function:
Shows the current display function.
•
Ambient Light Compensation:
Shows the ambient light compensation status.
•
Reading room:
Shows the selected reading room.
To retrieve the DICOM status report:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM Status Report
to make the information visible on the screen.
4.18.3 DICOM compliance check
About DICOM compliance check
The DICOM compliance check will measure the DICOM curve of your display in different steps. After
measurement, the DICOM status report is shown.
To start DICOM compliance check:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM Compliance Check
to start the compliance check.
Warning:
Pressing a key during the compliance check will abort the check.
Summary of Contents for Nio Fusion 12MP
Page 1: ...ENABLING BRIGHT OUTCOMES MDNC 12130 User guide Nio Fusion 12MP...
Page 10: ...R5913350 05 Nio Fusion 12MP 10 Welcome...
Page 20: ...R5913350 05 Nio Fusion 12MP 20 Installation and setup...
Page 26: ...R5913350 05 Nio Fusion 12MP 26 Advanced operation...
Page 42: ...R5913350 05 Nio Fusion 12MP 42 Advanced operation...
Page 43: ...43 R5913350 05 Nio Fusion 12MP Cleaning the display 5...
Page 45: ...45 R5913350 05 Nio Fusion 12MP Repackaging instructions 6...
Page 49: ...49 R5913350 05 Nio Fusion 12MP Image 6 7 Repackaging instructions...
Page 50: ...R5913350 05 Nio Fusion 12MP 50 Repackaging instructions...
Page 68: ...R5913350 05 Nio Fusion 12MP 68 Important information...
Page 69: ......
Page 70: ...www barco com R5913350 05 2022 12 19 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...