17
R5914634 /01 MDSC-8527
Protective earth
Earth the MDSC-8527 by connecting the protective earth pin to a grounded outlet by means of a yellow/green
AWG18 wire (maximum admitted cable length according to national regulation requirements).
Image 2–5
CAUTION:
The display must be earthed.
Potential equalization
When potential equalization between the display and other devices is required then connect the potential
equalization pin (POAG) to the potential equalization terminal of the equipment.
Image 2–6
2.4 Cable routing
To route the cables
Route all cables through the cable routing cutout while reinstalling the connector compartment.
Image 2–7
CAUTION:
When the display is assembled in the medical system, take care of the fixation of all
cables, to avoid unwanted detachment.
Summary of Contents for MDSC-8527
Page 6: ...R5914634 01 MDSC 8527 6 Welcome ...
Page 12: ...R5914634 01 MDSC 8527 12 Welcome ...
Page 13: ...13 R5914634 01 MDSC 8527 Display installation 2 ...
Page 19: ...19 R5914634 01 MDSC 8527 Daily operation 3 ...
Page 39: ...39 R5914634 01 MDSC 8527 Troubleshooting 5 ...
Page 41: ...41 R5914634 01 MDSC 8527 Important information 6 ...
Page 62: ...R5914634 01 MDSC 8527 62 Important information ...
Page 63: ......
Page 64: ...www barco com R5914634 01 2022 03 17 Barco NV President Kennedypark 35 8500 Kortrijk Belgium ...