5. Advanced operation
To enable the out power
1. Bring up the OSD main menu.
2. Navigate to the
Input Channel
menu.
3. Enter the
Input Channel
menu.
4. Navigate to the
Out Power
submenu.
5. Enter the
Out Power
submenu.
6. Enable or disable the Out power as desired and con
fi
rm.
5.3
Out Mode menu
Overview
•
5.3.1
Transfer function
About transfer function
There are different transfer function possibilities:
•
Native: the transfer function is transparent (no correction is applied).
•
DICOM (Factory): The output curve follows the DICOM curve and an offset can be applied to the curve
(see "User setting (DICOM)", page 23).
•
DICOM (User): The output curve follows the DICOM curve and an offset can be applied to the curve
(see "User setting (DICOM)", page 23).
The user curves are pre-setted in the factory or downloaded from User.
•
COLOR (Factory): The monitor output is corrected according to the color temperature pre-setted in
the factory. This setting can be set from 2000°K to 12000°K in steps of 100°K) (see "User setting
(COLOR)", page 23).
•
COLOR (User): The monitor output is corrected according to the color temperature selected by the
user. This setting can be set from 2000°K to 12000°K in steps of 100°K) (see "User setting (COLOR)",
page 23).
Monitor calibration is pre-setted in the factory but can be downloaded from User.
•
User LUT: The monitor output is corrected according to a prede
fi
ned user LUT.
The user curves are pre-setted in the factory or downloaded from User.
•
RGB Set (User): The monitor output is corrected according to the Red, Green and Blue color values
adjusted by the end user (from 0 to 1785 in steps of 1) (see "User setting (RGB SET)", page 24).
In this menu, COLOR is replaced with the current color temperature (e.g. 6500°K (FAC-
TORY))
To change the transfer function
1. Bring up the OSD main menu.
2. Navigate to the
Out Mode
menu.
3. Enter the
Out Mode
submenu.
4. Select the desired setting and con
fi
rm.
22