4.14 Ambient Light Compensation (ALC)
About ALC
Ambient Light Compensation (ALC) can only be enabled on your display when the DICOM display
function is selected. Therefore, please refer to
“
”
, page 28 to correctly set the
display function.
When ALC is enabled, the DICOM display function will be recalculated taking a preset ambient light correction
value into account. This value is determined by the selected reading room. Therefore, it is also important to
select a realistic reading room when enabling ALC. This can be done by following the instructions in
, page 29.
To enable/disable ALC:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Ambient Light
menu.
3.
Enter the
Ambient Light Compensation
submenu.
4.
Select
Enabled/Disabled
as desired and confirm.
4.15 Reading rooms
About reading rooms
Reading rooms can only be selected when the DICOM display function is selected. Therefore,
please refer to
“
”
, page 28 to correctly set the display function.
The American Association of Physicists in Medicine (AAPM) composed a list of pre-defined reading rooms.
Each of these reading rooms are defined by following parameters:
•
the maximum light allowed in this type of room
•
the preset ambient light correction value for this reading room
These parameters are stored in your display and determine the preset ambient light correction value to take
into account to recalculate the DICOM display function when Ambient Light Compensation (ALC) is enabled.
Please refer to
Ambient Light Compensation (ALC)
, page 29 to enable ALC.
The available reading rooms for your Coronis Fusion 4MP / 6MP are:
•
CR/DR/ MAMMO:
Corresponds to light conditions in diagnostic reading rooms for computed radiology,
digital radiology or mammography. This setting has the lowest maximum ambient light.
•
CT/MR/NM:
Corresponds to light conditions in diagnostic reading rooms for computed tomography,
magnetic resonance or nuclear medicine scans.
•
Staff Office:
Corresponds to light conditions in office rooms.
•
Clinical Viewing Room:
Corresponds to light conditions in diagnostic reading rooms for clinical viewing.
•
Emergency Room:
Corresponds to light conditions in emergency rooms.
•
Operating Room:
Corresponds to light conditions in operating rooms. This setting has the highest
maximum ambient light.
To select a reading room:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Ambient Light
menu.
3.
Enter the
Reading Room
submenu.
Summary of Contents for MDCC-4330
Page 1: ...ENABLING BRIGHT OUTCOMES MDCC 4330 MDCC 6430 User Guide Coronis Fusion 4MP 6MP...
Page 5: ...5 K5902108 06 Coronis Fusion 4MP 6MP Welcome 1...
Page 8: ...K5902108 06 Coronis Fusion 4MP 6MP 8 Welcome...
Page 18: ...K5902108 06 Coronis Fusion 4MP 6MP 18 Display installation...
Page 19: ...19 K5902108 06 Coronis Fusion 4MP 6MP Daily operation 3...
Page 23: ...23 K5902108 06 Coronis Fusion 4MP 6MP Advanced operation 4...
Page 36: ...K5902108 06 Coronis Fusion 4MP 6MP 36 Advanced operation...
Page 37: ...37 K5902108 06 Coronis Fusion 4MP 6MP Cleaning your display 5...
Page 39: ...39 K5902108 06 Coronis Fusion 4MP 6MP Repackaging instructions 6...
Page 42: ...K5902108 06 Coronis Fusion 4MP 6MP 42 Repackaging instructions...
Page 43: ...43 K5902108 06 Coronis Fusion 4MP 6MP Important information 7...