5. Advanced operation
5.10 Self calibration frequency
About self calibration
The front sensor of your Eonis display measures the output luminance of your screen and allows the
display to automatically stabilize its luminance for maximum image quality over the displays’ lifetime. This
self calibration is done at an adjustable, prede
fi
ned frequency:
•
1 min
•
1 hr
•
6 hr
•
24 hr
•
Never
•
QAWeb
: This setting will be automatically selected when the self calibration frequency is de
fi
ned by
MediCal QAWeb.
To adjust the self calibration frequency
1. Bring up the OSD main menu.
2. Navigate to the
Adjustments
>
Settings
menu.
3. Enter the
Self Calibration Frequency
submenu.
4. Select one of the available frequency presets.
5.11 Factory reset
About factory reset
A factory reset allows you to fully restore the display to its original factory setting.
To perform a factory reset
1. Bring up the OSD main menu.
2. Navigate to the
Adjustments
>
Settings
menu.
3. Enter the
Factory Reset
submenu.
4. Select
Yes
or
No
as desired and con
fi
rm.
K5903081 EONIS 19/09/2014
19