7. Important information
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment –
guidance
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
< 5% U
T
1
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in
U
T
) for 5s
< 5% U
T
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in
U
T
) for 5s
Mains power quality
should by that of a typical
commercial or hospital
environment. If the user
of the Eonis display
requires continued
operation during power
mains interruptions, it is
recommended that the
Eonis display be powered
from an uninterruptible
power supply or a battery.
1. is the a.c. mains voltage prior to application of the test level.
2. Eonis display doesn’t contain susceptible components to magnetic fields
K5903049 EONIS 11/10/2016
29