cause the device or a battery to explode. Avoid leaving your device under prolonged high temperatures such
as those above 45 °C / 113 °F.
The Demetra Scope requires adequate ventilation during proper use. Always ensure that the Demetra Scope
has adequate ventilation and air flow. Covering the device can significantly affect air flow, and may affect the
performance of the device and poses a possible risk of explosion or fire, which could lead to serious injuries or
damage to property.
Transient over-voltage
If the device is not used for a long time, remove it from the battery charging station to avoid damage by
transient over-voltage.
To fully disengage the power to the device, please remove it from the battery charging station.
Image review
To have sufficient image quality, a suitable display should be used to review images in the Demetra Web
Application (see the part
‘
Requirements
’
in the chapter
“
”
, page 12
for more details).
High magnetic environment
•
The device shall not be used in the high magnetic environment of an MRI scanner.
•
The installer shall assess the magnetic environment before installation or use of the device.
Water and moisture
Never expose the device to rain or moisture.
Never use the device near water - e.g. near a bathtub, washbasin, swimming pool, kitchen sink, laundry tub or
in a wet basement.
Eye safety
Hold your Demetra Scope so that the LEDs are aimed in the direction of the lesion to be examined.
Do not look directly into the bright LED light. It is recommended that patients close their eyes during
examination near their eyes.
The Demetra Scope is not intended to be used for lesions in the eyes.
Do not stare at operating lamp. May be harmful to the eyes. Product tested against IEC62471.
Glass temperature
The temperature of the contact glass can reach 46.5 °C.
Please verify the temperature of the contact glass and the condition of the patient before using contact mode.
Limit the contact duration to less than 1 minute, or use non-contact mode instead.
Breached or compromised surface
Do not make contact between the Demetra Scope and breached or compromised surfaces.
Risk patients
As the Demetra Scope contains several magnets (in the non-contact cone and in the battery charging station),
pay attention when examining specific risk patients (e.g. patients with a pacemaker, ...).
For those patients, it is recommended to use the device without the non-contact cone attached.
Malfunctions
Contact the Barco helpdesk under the following conditions:
•
If the power cord or plug is damaged or frayed.
•
If liquid has been spilled into the equipment.
•
If the equipment does not operate normally when the operating instructions are followed.
•
If the equipment has been dropped or has been damaged.
•
If the equipment exhibits a distinct change in performance, indicating a need for service.
Summary of Contents for Demetra
Page 1: ...ENABLING BRIGHT OUTCOMES BDEM 01 User Guide Barco Demetra...
Page 10: ...K5902200 07 Barco Demetra 10 Welcome...
Page 42: ...K5902200 07 Barco Demetra 42 Troubleshooting...
Page 43: ...43 K5902200 07 Barco Demetra 6 1 Repacking overview 44 Repacking instructions 6...
Page 45: ...45 K5902200 07 Barco Demetra 11 Plugs Repacking instructions...
Page 46: ...K5902200 07 Barco Demetra 46 Repacking instructions...
Page 47: ...47 K5902200 07 Barco Demetra 7 1 Instructions 48 Cleaning and disinfection 7...
Page 66: ...K5902200 07 Barco Demetra 66 Important information...
Page 67: ......