BANTE320 PORTABLE PH/ION METER 

28 

 
 

Specifications 

 

MODEL 

BANTE320 

pH 
Range 

-2.000~20.000pH 

Accuracy 

0.002pH 

Resolution 

0.1, 0.01, 0.001pH 

Calibration Points 

1 to 5 points, USA, NIST, DIN Standards 

Temperature Compensation 

0~100°C, 32~212°F, Manual or Automatic 

mV 
Range 

-1999.9~1999.9mV 

Accuracy 

± 0.2mV 

Resolution 

0.1, 1mV 

Calibration Points 

1 point (Only for Relative mV mode) 

Calibration Range 

± 200mV 

Ion 
Range 

0.001~19999ppm, mg/L, mol/L (Depending on range of ISE) 

Accuracy 

± 0.5% F.S (Monovalent), ± 1% F.S (Divalent) 

Resolution 

0.001, 0.01, 0.1, 1 

Calibration Points 

2 to 5 points 

Calibration Solutions 

0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000ppm, mol/L, mg/L 

Others 
Hold Function 

Manual or Automatic 

Stability Conditions 

Low or High 

Calibration Due 

0 to 31 days 

Power Off 

Manual or Automatic (10, 20, 30 minutes) 

Reset Function 

Yes 

Memory 

Stores up to 500 data sets 

Output 

USB Communication Interface 

Connector 

BNC 

Display 

LCD 

Power Requirements 

3×1.5V “AA” Batteries 

Dimensions 

170(L)× 85(W)× 30(H)mm 

Weight 

300g 

 
 
 
 

Summary of Contents for Bante320

Page 1: ...Bante320 Portable pH ORP Ion Meter Instruction Manual BANTE INSTRUMENTS LIMITED Technology to create the future...

Page 2: ...a clear and bright LCD display that used to show measured values mode indicators and message indicators INDEX Measurement mode indicator Indicates meter is in the measurement mode Low Battery Alarm Wh...

Page 3: ...CAL Press the key to enter the calibration mode Press and hold the key to enter the setup menu MI Press the key to store current measured value Press key in setup mode to scroll up through menu Press...

Page 4: ...rt the batteries into the battery compartment note polarity 3 Replace the battery cover into its original position Installation is completed When batteries are depleted the meter allows you to use the...

Page 5: ...three pH buffer packets required for calibration Open the pH7 00 buffer packet place the powder into a 250ml volumetric flask Pour distilled water 250ml to scale line mix the solution until reagent i...

Page 6: ...s Select the number of calibration points 1 point 2 points 3 points 4 points 5 points Resolution Sets the resolution for pH measurement 0 001pH 0 01pH 0 1pH Measurement Unit Sets the default temperatu...

Page 7: ...ense a stable endpoint reading and freeze it Enable Disable Auto Power Off When the auto off is enabled if you do not press any key within a specified time period the meter will automatically turn off...

Page 8: ...electable parameter and page number 3 Press or key to scroll through menu select the parameter you want to set Refer to Setup Menu section 4 Press ENTER key to confirm the display shows an option in t...

Page 9: ...til the meter shows Date 3 Press ENTER key to confirm the display shows current year 4 Press or key to set the year 5 Press ENTER key to confirm the meter shows current date and time 6 Press or key to...

Page 10: ...lay the meter is now switched to automatic temperature compensation mode MANUAL TEMPERATURE COMPENSATION 1 DO NOT connect the temperature probe to meter 2 Press and hold the o C key for 3 seconds to e...

Page 11: ...cepted Make sure to calibrate the meter when attaching a new electrode or during first use DO NOT reuse calibration solution after calibration contaminants in solution will affect the calibration and...

Page 12: ...2 3 Rinse the pH electrode with distilled water Dip the electrode into the next standard buffer solution e g pH4 01 The meter automatically sense the current calibration solution and begin calibratin...

Page 13: ...econd point calibration 3 5 Rinse the pH electrode with distilled water Dip the electrode into the next custom buffer solution 3 6 Press or key to increase or decrease displayed value press ENTER key...

Page 14: ...ress or key until the display shows ELE Electrode Diagnosis 3 Press ENTER key to confirm the meter shows the last calibration date Format mm dd 4 Press key the meter shows OFS Zero point offset potent...

Page 15: ...DE key until the meter shows indicator 2 Rinse the ORP electrode with distilled water Dip the electrode into the calibration solution Stir the sensor gently 3 Press CAL key the meter shows current mea...

Page 16: ...s MR key in the ORP measurement mode the meter shows LOC 2 Press or key until the display shows ELE Electrode Diagnosis 3 Press ENTER key to confirm the meter shows the last calibration date Format mm...

Page 17: ...nt results we recommend that you perform ion calibration and measurement at same temperature 1 Press MODE key until the meter shows indicator 2 If necessary select the concentration unit and ion valen...

Page 18: ...to stabilize the display shows 0 1 CAL2 The meter prompts you to continue with second point calibration 7 Dip the Ion Selective Electrode into the corresponding calibration solution Stir the sensor g...

Page 19: ...easurement mode the meter shows LOC 2 Press or key until the display shows ELE Electrode Diagnosis 3 Press ENTER key to confirm the meter shows the last calibration date Format mm dd 4 Press key the m...

Page 20: ...r key to set the temperature value 3 Press ENTER key to confirm Calibration is completed pH Measurement 1 Press MODE key until the meter shows indicator 2 Rinse the pH electrode thoroughly with distil...

Page 21: ...ve Electrode thoroughly with distilled or de ionised water 3 Dip the electrode into the sample solution stir the sensor gently 4 Wait for the reading to stabilize record the measured value on the disp...

Page 22: ...indicator appears on the display MEMORY RECALL 1 Press MR key in the measurement mode the meter shows LOC Data Log 2 Press ENTER key to confirm the meter shows page number of the stored data set 3 Pre...

Page 23: ...n on the start bar the DAS splash screen appears for few seconds and then the application starts RECEIVING THE STORED DATA Connect the USB cable to meter and computer Communication port COM1 Click the...

Page 24: ...need to view previous data click the Recall button the screen will show all data in the database DELETE THE DATA If you need to delete the certain data click on the data field then click the Delete b...

Page 25: ...een will show a dialog box Selecting the time interval and recording time click the OK button the measured value will automatically transfer to data sheet CREATE THE EXCEL DATA FILE When the transfer...

Page 26: ...n for 10 minutes Rinse the sensor with distilled water REACTIVATING THE ELECTRODE If stored and cleaned properly the electrode should be ready for immediate use However a dehydrated bulb may cause slu...

Page 27: ...ement requirements please refer to the table below to select an applicable sensor Sample Type P11 P12 P13 P14 P16 P17 P18 P19 P20 P21 Agar Beer Blood Products Bread Dough Cement Cosmetics Dairy Produc...

Page 28: ...Iodide Pb US Lead K US Potassium Ag US Silver S US Sulphide Troubleshooting LCD DISPLAY CAUSE CORRECTIVE ACTION Electrode dried out Soak the pH electrode in 3M KCL solution for 10 minutes Soak the Io...

Page 29: ...mV Ion Range 0 001 19999ppm mg L mol L Depending on range of ISE Accuracy 0 5 F S Monovalent 1 F S Divalent Resolution 0 001 0 01 0 1 1 Calibration Points 2 to 5 points Calibration Solutions 0 001 0 0...

Page 30: ...hese efforts and we expect to have little or no such materials in our product in the coming years We welcome our customer suggestions on how to speed up these efforts Warranty The warranty period for...

Reviews: