Bante900P Portable Multiparameter Water Quality Meter
24
8. Rinse the Ion Selective Electrode with distilled water. Dip the electrode into corresponding calibration solution. Stir the sensor gently.
9. Press ENTER key, “Calibration” indicator begins flashing. Wait for the measured value to stabilize, the display shows “1/CAL3”. The meter prompts you to
continue with third point calibration.
10. Repeat steps 8 to 9 above until the display shows “END”, the meter returns to measurement mode automatically. Calibration is completed.
EXIT THE CALIBRATION:
During the calibration process, if you want to exit calibration, press MEAS key, the meter will return to measurement mode immediately.
Enter