Specifications

 

 

Specifications 

 

Model 

TB100 

Measurement Method 

ISO 7027 nephelometric method (90° ) 

Range 

0

 

~

 

1100 NTU/FNU, 0

 

~

 

275 EBC, 0

 

~

 

9999 ASBC 

Resolution 

0.01 (0

 

~

 

100 NTU), 0.1 (100

 

~

 

999 NTU), 1 (999

 

~

 

1100 NTU) 

Accuracy 

± 2% (0

 

~

 

500 NTU), ± 3% (501

 

~

 

1100 NTU) 

Calibration Standards 

0.02, 10.00, 200, 500, 1000 NTU 

Light Source 

Infrared-emitting diode 

Detector 

Silicon photodiode 

Stray Light 

< 0.02 NTU 

Sample Vial 

60

 

(H) ×  25

 

(Dia)

 

mm 

Memory 

Stores up to 100 data sets 

Output 

USB communication interface 

Display 

Segmented LCD, 60

 

(L) ×  40

 

(W)

 

mm 

Operating Temperature 

0

 

~

 

50° C, 32

 

~

 

122° F 

Power Requirements 

3 ×  1.5V AA alkaline batteries or DC5V power adapter 

Dimensions 

180

 

(L) ×  85

 

(W) ×  70

 

(H)

 

mm 

Weight 

300g 

 

Optional Accessories 

 

Order Code 

Description 

TB-GV 

Sample vial, glass body, 60

 

(H) ×  25

 

(Dia)

 

mm 

USB-AC 

USB cable, 1 meter, USB-A plug 

DCPA-5V 

DC5V power adapter, 1 meter cable, european standard plug 

 
 
 
 

 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for TB100

Page 1: ...TB100 Portable Turbidity Meter Instruction Manual Bante Instruments Inc...

Page 2: ...ber of calibration points 7 Calibrating the meter 8 Turbidity measurement 8 Selecting the measurement mode 8 Single measurement low turbidity samples 200 NTU 9 Single measurement high turbidity sample...

Page 3: ...of meter After the unpacking please check all components are complete If any are damaged or missing please contact the supplier The sample vial in the measurement chamber has been indexed with the 0...

Page 4: ...ment mode Starts calibration Enters the setup menu Press and hold the key for 3 seconds Stores current reading to memory Increase value or scroll up through the menu item Views the data log Decrease v...

Page 5: ...meter is in the calibration mode Calibration icon Indicates the meter is calibrating Installing the Batteries 1 Remove the battery compartment cover from the backside of meter 2 Insert the three AA ba...

Page 6: ...fault turbidity unit NTU Nephelometric turbidity unit NTU FNU Formazin nephelometric unit EBC Turbidity unit of the European Brewery Commission ASBC Turbidity unit of the American Society of Brewing C...

Page 7: ...e factory settings The Factory Reset option is used to restore the meter back to default settings If enabled all of the calibration data and selected options will be lost or reset the meter must be re...

Page 8: ...eaned thoroughly with deionised water after each measurement Any residues can cause inaccurate readings Turbidity Calibration The TB100 portable turbidity meter is calibrated with Formazin Standards a...

Page 9: ...ext calibration point 2 7 If necessary press the key to select an applicable calibration point 2 8 Repeat steps 2 2 through 2 5 until the meter shows Calibration is completed To exit the calibration m...

Page 10: ...flashing 1 6 Wait for the reading to stabilize the Measure icon will stop flashing Press the Meas key the meter will take a measurement again Single measurement high turbidity samples 1100 NTU For the...

Page 11: ...e key the display shows the date and time of measurement Format mm dd hh mm 3 Press the key the display shows the stored data 4 Press the key again the display shows the next data set 5 Press the Meas...

Page 12: ...computer 2 Click the OK button the application starts If your computer can not find a communication port please click the file CP210xVCPInstaller_x64 exe or CP210xVCPInstaller_x86 exe to update the d...

Page 13: ...f these vials measuring value and mark it Preparing the formazin standards Preparation of turbidity free water The turbidity free water is used for preparation of turbidity standards and is prepared b...

Page 14: ...Detector Silicon photodiode Stray Light 0 02 NTU Sample Vial 60 H 25 Dia mm Memory Stores up to 100 data sets Output USB communication interface Display Segmented LCD 60 L 40 W mm Operating Temperatu...

Page 15: ...r suggestions on how to speed up these efforts Warranty The warranty period for meter is two years from the date of shipment Above warranty does not cover the calibration standards Out of warranty pro...

Reviews: