POL-200 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Introduction 

Thank you for selecting the POL-200 semiautomatic polarimeter. This 
user manual provides a step-by-step guide to help you operate the   
instrument, please carefully read the following instructions before   
use. Any use outside of these instructions may invalidate your warranty 
and cause permanent damage to the instrument. 
 

Environmental Conditions 

Before unpacking, ensure that current environmental conditions meet 
the following requirements. 

  Relative humidity is less than 80% 

  Ambient temperature between 0° C (32° F) and 40° C (104° F) 

  No ambient light interference 

  No corrosive gas exists 

 

Packing List 

The following list describes all components of the instrument. If any 
items are missing or damaged, contact the supplier immediately. 
 

 

 

  POL-200 polarimeter  

 

 

 

    Sealing rings 

 

 

  Test tubes 100 and 200 mm 

 

 

 

Switching the Instrument On and Off 

  Plug the power cord into the wall outlet. Press the 

ON/OFF

 key, 

 

the screen shows "

Wait for light source to stabilize

 

". After a few 

 

seconds, the instrument shows default measurement screen and 

 

ready for use. 

  Press the 

ON/OFF

 key again, the instrument begins running and 

 

returns to the optical zero point, the screen will automatically   

 

switch off. 

 

 

WARNING 

 

Do not pull out the power cord if the instrument is not completely 

 

switched off, that will cause the optical and mechanical faults. 

 

 

 

Display 

The POL-200 polarimeter is installed with the 5.6 inch touch screen for 
showing the function key, measured value, associated parameters and 
operation guide. In the measurement and setting modes, the screen 
will show different user interfaces. 
 

 

 

 

 

 

 

 

Settings 

The instrument contains an integrated setup menu for customizing the 
function parameters to meet measurement requirements, selectable 
items include the following options. 
 

 

 

 

 

 

Setup Menu 

Measurement Mode 

Optical Rotation (default) 
Specific Rotation 
Concentration 
International Sugar Scale 

Tube Length 

100 mm (default) 
200 mm 
Custom (range: 0.0 to 220.0 mm) 
 

 

Function   

keys 

Guide 

Parameter 

Summary of Contents for POL-200

Page 1: ...POL 200 Semiautomatic Polarimeter USER MANUAL Bante Instruments Inc ...

Page 2: ...all outlet Press the ON OFF key the screen shows Wait for light source to stabilize After a few seconds the instrument shows default measurement screen and ready for use Press the ON OFF key again the instrument begins running and returns to the optical zero point the screen will automatically switch off WARNING Do not pull out the power cord if the instrument is not completely switched off that w...

Page 3: ...ey at the top of the screen and select an option 2 Tap the or key to select a parameter or the numeric keys to enter the setting value 3 Tap the Enter to confirm Prior to Use Rotate the focus knob on front panel of the instrument until the visual field becomes legible Zero Point Correction 1 Place an empty test tube into the sample chamber 2 If the visual field appears the different brightness tap...

Page 4: ...e instrument stops running Measurement is completed During the measurement press and hold the or key for about 5 seconds the instrument will switch to the auto run mode Tap the or key again the instrument will switch to the manual mode Data Management The POL 200 polarimeter is capable of storing and recalling up to 100 data sets Storing Data into Memory 1 1 In the measurement process tap the Save...

Page 5: ...orm brightness Tap the Enter key the instrument will automatically returns to the optical zero point and stops running If you have selected multipoint calibration in setup menu repeat the steps 5 through 6 until the screen shows Calibration is completed If 1 point calibration is selected the instrument is only allowed zero point calibration Appendix Optional Accessories Order Code Description POLT...

Page 6: ...warranty period for semiautomatic polarimeter is one year from the date of shipment Out of warranty products will be repaired on a charged basis The warranty on your polarimeter shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products For more information please c...

Reviews: