PHscan30L 

 

Switching the Tester On and Off 

 

Press and hold the 

Meas

 key for about 5 seconds to switch on the tester. 

 

Press and hold the 

Meas

 key to switch off the tester. 

 

To enable the auto-power off function, refer to the 

Setup Menu

 section.

 

 

Setup Menu 

The PHscan30 tester contains 7 menu items in the setup menu, the following table describes the functions of each option. 
 

Menu 

Description 

Options 

Description 

Default 

 

Set the pH buffer group for calibration and auto-recognition 

 

USA (pH

 

4.01, 7.00, 10.01) 

USA 

 

NIST (pH

 

4.01, 6.86, 9.18) 

 

Set the number of calibration points 

 

1, 2 or 3 points 

2 points 

 
 

 

Set the temperature unit 

 

Degrees Celsius 

° C 

 

Degrees Fahrenheit 

 

Calibrate the temperature 

 

Reading ± 10° C/° F 

--- 

 

 

When the option is enabled, the tester will automatically   
sense a stable reading and lock the measurement 

 

Enable 

Disable 

 

Disable 

 

When the option is enabled, the tester will automatically   
switch off if no key is pressed within 8 minutes 

 

Enable 

Enable 

 

Disable 

 

Reset the tester to factory default settings 

 

Enable 

Disable 

 

Disable 

 

Setting the Default Option 

1. 

In the measurement mode, press and hold the 

Cal

 key for 5 seconds to enter the setup menu. 

2. 

If necessary, press the 

Cal

 key again to select an option. 

3. 

Press the 

Enter

 key, the tester saves the current option and moves to the next menu item. 

4. 

Repeat the steps above until the tester returns to the measurement mode. Setting is completed. 

 

 

 

 

 

 

During the setting process, press the 

Meas

 key, the tester will exit the setup menu and return to the measurement mode. 

 

If you do not need to calibrate the temperature, press the 

Enter

 key to skip the

 

/

 

 

or

 

/

 

 

option. 

 

The

 

 

option is used to restore the tester back to the factory default settings. If enabled, all of the calibration data and user-specific 

 

settings will be deleted or reset, the tester must be recalibrated. 

option 

menu 
item 

Summary of Contents for PHscan30L

Page 1: ...PHscan30L Pocket pH Tester Instruction Manual Bante Instruments Inc...

Page 2: ...e pH measurement mode Start calibration Press and hold key to enter the setup menu Select an option Confirm the calibration setting or displayed option Display Icon Description When the battery voltag...

Page 3: ...Disable Disable When the option is enabled the tester will automatically switch off if no key is pressed within 8 minutes Enable Enable Disable Reset the tester to factory default settings Enable Disa...

Page 4: ...tester shows pH7 00 CAL 1 or pH6 86 CAL 1 depending on the selected pH buffer group 1 2 Rinse the electrode with deionized water and place into the pH7 00 buffer solution Stir the tester gently to cre...

Page 5: ...Temperature Calibration The PHscan30 tester has a built in temperature sensor that used for automatic temperature compensation ATC Durning the measurement if the measured temperature reading differs f...

Page 6: ...into a 250 ml volumetric flask Fill the deionized water to the mark and mix the solution until the reagent is completely dissolved Preparation of pH4 01 and 10 01 buffer solutions are the same as abov...

Page 7: ...racy 1 C 1 8 F Calibration Point 1 point reading 10 C Other Specifications Operating Temperature 0 to 50 C 32 to 122 F Storage Temperature 5 to 60 C 23 to 140 F Relative Humidity 80 non condensing Dis...

Page 8: ...ce 4715 Castlewood St Sugar land TX 77479 USA Tel 1 346 762 7358 E mail banteinstruments yahoo com Factory A10 No 2066 Laifang Rd Shanghai 201615 China Tel 86 21 6404 1598 E mail banteinstrument hotma...

Reviews: