![Bante Instruments ECscan10 User Manual Download Page 4](http://html1.mh-extra.com/html/bante-instruments/ecscan10/ecscan10_user-manual_3727671004.webp)
3
I
ECscan10
Electrode
Maintenance and Replacement
•
Rinse the electrode thoroughly with distilled water after use.
•
Do not touch the platinum black coating on the sensor surface and
always keep it clean.
•
If there is a build-up of solids inside the sensor, remove carefully,
then recalibrate the tester.
•
If you do not use the tester for long periods, store the electrode
with tap water.
Replacing the Electrode
If the tester fails to calibrate or gives fluctuating readings, you should
consider replacing the electrode.
1. Twist the electrode collar counter clockwise, pull the electrode
away from the tester.
2. Align the slot on the new electrode, gently push the electrode into
the tester.
3. Twist the electrode collar clockwise until tight.
Appendix
Preparation of Conductivity Standard Solutions
Place the analytical grade potassium chloride (KCl) in a beaker and
dry in an oven for about 3 hours at 105° C (221° F), then cool to room
temperature. Add the reagent to a 1 liter volumetric flask according
to the instructions in table below. Fill the distilled water to the mark,
mix the solution until the reagent is completely dissolved.
Conductivity Standard
Reagent
Weight
146.5 µ S/cm
KCl
74.4 mg
1413 µ S/cm
KCl
745.5 mg
12.88 mS/cm
KCl
7.45 g
Optional Accessories
Order Code
Description
E-ECscan-C1-100K
Platinum conductivity cell, K=1
ECCS-146
Standard solution 146.5
μ
S/cm, 480
ml
ECCS-1413
Standard solution 1413
μ
S/cm, 480
ml
ECCS-1288
Standard solution 12.88
mS/cm, 480
ml
Tester Specifications
Model
ECscan10
Conductivity
Range
ECscan10
L: 1.0 to 199.9
μ
S/cm
ECscan10
M: 10 to 1999
μ
S/cm
ECscan10
H: 0.1 to 19.99
mS/cm
Resolution
0.1, 1
Accuracy
± 1% F.S
Calibration Point
1 point
Calibration Solution
ECscan10
L: 146.5
μ
S/cm
ECscan10
M: 1413
μ
S/cm
ECscan10
H: 12.88
mS/cm
Temperature Compensation 0 to 50° C, automatic
Temperature
Range
0 to 50° C
Resolution
1° C
Accuracy
± 1° C
Other Specifications
Cell Constant
K=1
Operating Temperature
0 to 50° C (32 to 122° F)
Storage Temperature
0 to 60° C (32 to 140° F)
Relative Humidity
< 80% (non-condensing)
IP Rating
IP54
Display
LCD, 21 × 21 mm (0.82 × 0.82 in.)
Power Requirements
3 × 1.5V LR44 micro alkaline batteries
Auto-Off
8 minutes after last key pressed
Dimensions
185 (L) × 40 (Dia.) mm (7.28 × 1.57 in.)
Weight
100