3   

I   

Overview 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Connecting the Temperature Probe

 

2.1  Place the temperature probe into the circular hole located at the 
 

center of the electrode arm. 

 

 

 

 

 

 
2.2  Insert the jack plug to the connector socket labeled ATC. Ensure 
 

the connector is fully seated. 

 

 

 

Connecting the Power Adapter 

3.1  Insert the connector of power adapter to the power socket. 
3.2  Plug the 9V DC power adapter into the wall outlet. 
 

 

 

 

 

Prior to Use 

Soak the conductivity electrode for about 10 minutes in tap water to 
remove dirt and oil stains on the sensor surface. 
 

10

 minutes 

 

 

 
 
 

 

 

Switching the Meter On and Off 

  Press the 

Meas

 key for about 3 seconds and release to switch   

 

on the meter. 

  Press and hold the 

Meas

 key to switch off the meter. 

 

 

 

 

 

Meter Setup 

The 12DW meter contains an integrated setup menu for customizing 
the displayed option to meet measurement requirements. The   
following table describes the functions of each menu item. 

 
Menu Item 

Option and Description 

 

Cell Constant 

Set the cell constant to match connected conductivity 
electrode. 

 

K = 0.1 

 

K = 1 (default) 

 

K = 10 

 

Custom 

 

Calibration Points 

Set the number of calibration points. 

 

1 point (default) 

 

2 points 

 

3 points 

 

Temperature Coefficient 

Set the temperature coefficient for linear temperature 
compensation. 

 

0.0 to 9.9%/° C (default 2.1) 

 

Summary of Contents for DDS-12DW

Page 1: ...DDS 12DW Benchtop Conductivity Meter USER MANUAL Bante Instruments Inc...

Page 2: ...g List The following list describes all components of the meter If any items are missing or damaged contact the supplier immediately DDS 12DW meter Electrode arm 9V DC power adapter CON 1 conductivity...

Page 3: ...ectrode arm 90 degrees The electrode holder is now ready to swing into desired position Adjusting the Electrode Arm After installation if the electrode arm automatically rises or falls you are able to...

Page 4: ...e 10 minutes Switching the Meter On and Off Press the Meas key for about 3 seconds and release to switch on the meter Press and hold the Meas key to switch off the meter Meter Setup The 12DW meter con...

Page 5: ...0 the temperature compensation will be disabled and the meter only shows the actual conductivity at the measured temperature Automatic Temperature Compensation Connect the temperature probe to meter...

Page 6: ...e standard solution stir gently to remove air bubbles trapped in the slot of the sensor The meter will automatically show the calibration standard e g 1413 S cm 1 4 Press the Enter key the default cal...

Page 7: ...rs on the display Press the Meas key to resume measuring If the option is disabled the meter will continuously measure and update the readings 3 Wait for the measurement to stabilize and record the re...

Page 8: ...ture setting 3 Press the key to set the temperature to 25 C and press the Enter key to confirm 4 Place the conductivity electrode into the sample solution record the temperature value TA and conductiv...

Page 9: ...mperature 0 to 50 C 32 to 122 F Storage Temperature 0 to 60 C 32 to 140 F Relative Humidity 80 non condensing Display LCD 135 75 mm 5 3 2 9 in Power Requirements 9V 400mA DC power adapter Auto Off 3 h...

Page 10: ...eter is one year from the date of shipment Above warranty does not cover the electrode and standard solutions Out of warranty products will be repaired on a charged basis The warranty on your meter sh...