DOscan10 

 

Dissolved Oxygen Calibration 

The DOscan10 tester allows 1 or 2 points calibration. If you have selected the single point calibration in the setup menu, we recommend that you 
perform a 100% saturation calibration in the air-saturated water. If the 2 points calibration is selected, the zero oxygen solution needs to be used. 
During the calibration and measurement, the temperature sensor on electrode must be immersed in solution completely, the solution keeps 0.3

 

m/s of minimum flow rate to avoid oxygen starvation at the membrane. 
 

 

 
 

Single Point Calibration - 100% Saturation 

Make sure that you have selected 1 point calibration in the setup menu. 
1.1  Press the 

Cal

 key, the display shows "100%

 

/

 

CAL1". 

1.2  Hold the dissolved oxygen electrode in the air at 100% relative humidity or place the electrode into the air-saturated water for about 10 
 

minutes. Press the 

Enter

 key to begin the calibration. 

1.3  When the reading has stabilized, the tester will show "End" and return to the measurement mode. 
 

 

 

 

2 Point Calibration

 

Make sure that you have selected 2 points calibration in the setup menu. 
2.1  Repeat steps 1.1 through 1.2 above. When the first calibration point is completed, the display will show "0%

 

/

 

CAL2", the tester prompts 

 

you to continue with second point calibration. 

2.2  Place the dissolved oxygen electrode into the zero oxygen solution for about 10 minutes, press the 

Enter

 key to begin the calibration. 

2.3  When the reading has stabilized, the tester will show "End". Calibration is completed. 
 

 

 

 

 

 

temperature sensor

 

 

 

Performing a percentage saturation calibration will simultaneously calibrate 

 

the corresponding mg/L (or ppm) concentration value. 

 

To exit the calibration without saving calibrated values, press the 

Meas

 key.

 

Summary of Contents for D0scan10

Page 1: ...DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual Bante Instruments Inc...

Page 2: ...ess the key again to resume measurement Exit the calibration setting and return to the measurement mode Start calibration Press and hold the key to enter the setup menu Select an option Confirm the ca...

Page 3: ...to polarize Setup Menu The DOscan10 tester contains 8 menu items in the setup menu the following table describes the functions of each option Menu Description Options Description Default Set the sali...

Page 4: ...enabled all of the calibration data and user specific settings will be deleted or reset the tester must be recalibrated The following table describes the relationship between the altitude and barometr...

Page 5: ...0 relative humidity or place the electrode into the air saturated water for about 10 minutes Press the Enter key to begin the calibration 1 3 When the reading has stabilized the tester will show End a...

Page 6: ...meate the membrane via diffusion Their existence will influence the measurements If the sample contains solvent grease sulfide and alga the membrane will be damaged or eroded 1 Rinse the electrode wit...

Page 7: ...DO ES Electrolyte solution 480ml Specifications Dissolved Oxygen Model DOscan10 Range 0 0 to 20 0mg L Resolution 0 1mg L Accuracy 0 5mg L Saturation Range 0 0 to 200 0 Resolution 0 1 Accuracy 2 0 Othe...

Page 8: ...warranty does not cover the electrode and membrane cap Out of warranty products will be repaired on a charged basis The warranty on your tester shall not apply to defects resulting from Improper or in...

Reviews: