61279e GB/2018-03
11 / 25
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
Notes:
- The height of the over
fl
ow edge depends on the lining of the work surface (see layout
samples in chapter 4.3).
- The over
fl
ow edge may protrude a maximum of 2 mm wide and a maximum of 10 mm
deep into the tank, so that the basket mounts for the accessories are not covered.
Before installation, verify the table speci
fi
cations since there may be production-
related deviations/tolerances. (If necessary refer to the diagram in chapter 6.)
- The HF cable on the side cover may not be kinked - additional distance: 10 cm. See
enclosed drawing.
- For the tank cut-out, take into account the position of the TRISON control unit and,
if needed, the space requirement for the pivot mounted arm TRISON TL 4000, see
chapters 3, 6 and 8.
- Thanks to the inclined tank bottom, the tank is able to empty out almost completely.
4.2
Fixing the sealing tape
Apply the self-adhesive sealing tape that is
included in the scope of supply to the tank edge,
this facilitates subsequent servicing.
Degrease
the tank edge
thoroughly
and then
apply the sealing tape as shown in the image,
allowing the two ends to overlap.
Note:
Always set down the ultrasonic oscillating tank
carefully on the oscillator cover.
Summary of Contents for TRISON Series