24
6.3
Error analysis
SONOREX ultrasonic and rinsing baths are robustly constructed and designed for a
high level of reliability.
Nevertheless, the possibility of a malfunction due to a defective component can never
be fully discounted.
The following overview of possible sources of error should serve as an aid for error
detection and elimination.
Ultrasonic bath oscillates weakly, unevenly, or noise is too loud:
- Has
fl
uid been properly degassed?
- Is it overloaded with objects to be treated?
- Uneven noises (wobbling)
Treat for 30 min.
Remove a few parts.
No error - slightly adjust the
fi
ll level
of the
fl
uid.
Heating system defective?
The ultrasonic bath can be readily operated
without heating.
Slight erosion visible on the bottom of the bath?
Natural wear.
Ultrasonic bath OK.
Strong erosion marks appear at the
tank bottom
lead to loss of power
Any malfunctions must be communicated in writing to the manufacturer.
6.4
Customer service
If service is required, please contact your specialist dealer or Bandelin electronic to
order spare parts or before sending in defective devices.
6.4.1 Repairs and service
If errors or defects are ascertained as a result of the functional check, and if it is
impossible to rectify such errors, the ultrasonic bath may no longer be used. In such a
case, please contact the supplier or the manufacturer: