Notes - Inserting objects
• Fully remove air bubbles from cavities (e.g. blind holes).
Indirect sonication
Remove any air bubbles from underneath the
vessels.
Notes - Temperature and heating
• Warmed-up
fl
uids intensify the ultrasound effect. Experience has shown that the best results
are obtained with a bath temperature of 50 to 60 °C. With high temperatures, the effect of the
ultrasound cavitation decreases; however
/3
.
• In order to save time during use, the bath
fl
uid may be preheated during degassing.
• For an optimum bath temperature, observe the speci
fi
cations of the specimen manufacturer!
• To protect the electronic components inside the ultrasonic bath, the ultrasound output is
reduced upon reaching a critical temperature in order to inhibit a further increase in the interior
temperature.
• The
fl
uid in the oscillating tank may not exceed a maximum operating temperature of 100 °C.
• Prepare the disinfectant with cold water, operate the ultrasonic bath without heating and avoid
temperatures above 40 °C.