POCKET PHOTOMETERS

MW line

28-04-2016

190 M1 11 D

Instructions  manual

Instructions  manual

Pocket photometers

MW line

FREE CHLORINE MEASUREMENT (MW-10)

Turn the meter ON by pressing the button. After
all segments are displayed, 

“C.1”

“Add”

appears

with "Press" blinking, the meter is ready. 

Fill the test tube with 

10 mL

of unreacted sample

and  tap  it  with  the  cap.  Place  it  into  the  tube
holder and close the meter cap. 

Press  the  button.  When  the  display  shows

“Add”

“C.2”

with 

“Press”

blinking,  zero  is

calibrated. 

Remove  the  test  tube,  open  it  and  add  the
content of one 

DPD1

reagent test. Replace the

cap and shake gently for 20 seconds. Replace
the tube into the meter. 

Wait for 1 minute and then press and hold the
button until the timer is displayed on the LCD. 

The instrument directly displays the concentration
of free chlorine in ppm. The meter automatically
turns off after 10 seconds following the measu-
rement display.

PACK CONTENT

Each photometer is supplied with:

• 2 test tubes with cap
• 1 battery 1.5 V, type AAA
• 1 instruction manual

TECHNICAL FEATURES

Range: 

0.05 to 2.50 ppm

Resolution:

0.01 ppm

Accuracy: 

± 0,05 ppm ± 3% of reading (25°C)

EMC Deviation: 

± 0,01 ppm

Light source:

LED at 525 nm

Light detector:

Silicon photocell

Method: 

Adaptation of USEPA method 330.5. 
The reaction between chlorine and the DPD 
reagent causes a pink tint in the sample.

Temperature: 

+1...+50°C

Humidity: 

85% RH, non condensing

Power supply: 

1 battery 1.5 V type AAA

Auto switch-off:

After 2 min

Dimensions – Mass:  87,5 x 63,7 x 35,2 mm – 65 g

REPLACEMENT PARTS

TOTAL CHLORINE MEASUREMENT (MW-11)

Turn the meter ON by pressing the button. After
all segments are displayed, 

“C.1”

“Add”

appears

with "Press" blinking, the meter is ready. 

Fill the test tube with 

10 mL

of unreacted sample

and  tap  it  with  the  cap.  Place  it  into  the  tube
holder and close the meter cap. 

Press  the  button.  When  the  display  shows

“Add”

“C.2”

with 

“Press”

blinking,  zero  is

calibrated.

Remove  the  test  tube,  open  it  and  add  the
content of 

one DPD1 and one DPD3 

reagent

tests. Replace the cap and shake gently for 20
seconds. Replace the tube into the meter. 

Wait for 2 minutes and 30 seconds then press
and hold the button until the timer is displayed
on the LCD. 

The instrument directly displays the concentration
of total chlorine in ppm. The meter automatically
turns off after 10 seconds following the measu-

rement display.

Dust cover

Test tube (with cap)

Tube holder

LCD

Touch button