450 M1 02 B

11-10-2007

ABOUT Bamophox

ABOUT BAMOPHOX

VERSION 2.04

SERIAL N°

20451 05

CONSULTATION / MODIFICATION

CONSULTATION

MODIFICATION

CODE ? 0000

CODE ? 5105

TIME : 30 mn

ENTER

ENTER

ENTER

ENTER

L

ENTER

ENTER

ENTER

MENU

+

L

L

From this mode MODIFICATION it is easy to return back to measurement for testing the relay outputs
and regulation mode.

Once in modification mode,

reach measurement display and press ENTER

FORCED MEASURE

00,00 mg/l  +20°C

(one digit is flashing) Modify the value. Immediately the 
instrument acts within the configuration (thresholds, 
regulation, analog outputs ...).

When PID regulation is activated, the display shows the PID %

FORCED PID

00,00 mg/l  0,000%

(one digit is flashing) Modify the value. Immediately the 
instrument acts within the configuration.

To test the analog output (mA) on PID mode: the PID should be active and in MANUAL mode.

Press ENTER to cancel the test mode and to go back to the measurement mode.

L

L

ENTER

ENTER

ENTER

ENTER

ENTER

L

L

Last 4 digits (of serial number) are the key code to
access the MODIFICATION menu. 
When wrong code is entered, a message 

"ERROR"

appears during 3 seconds.

After 30 minutes, the display returns automatically to the
measurement mode. 

Summary of Contents for BAMOPHOX 436 M

Page 1: ...amo fr Fax 33 01 34 10 16 05 E mail info bamo fr Dissolved oxygen monitoring BAMOPHOX 450 11 10 2007 450 M1 02 B MES 450 02 1 INSTRUCTION MANUAL INSTRUCTION MANUAL B B B BA A A AM M M MO O O OP P P PH...

Page 2: ...3 WIRING 4 4 FRONT PANEL 6 PRESENTATION AND SCROLLING MENU 7 ABOUT BAMOPHOX 8 CONSULTATION MODIFICATION 8 MEASUREMENT PARAMETERS 9 ATMOSPHERIC PRESSURE 9 SALINITY 9 ADJUST THRESHOLD 1 10 ADJUST THRESH...

Page 3: ...in 3 A 125 V AC 3 A 30 V DC and 105 evaluated charge for 3 A 125 V AC ON OFF Regulation Pulse time 0 9999 sec High and low proportional bandwidth high and low dead zones PID Regulation Proportionality...

Page 4: ...Page 4 450 M1 02 B 11 10 2007 3 WIRING PANEL MOUNTING MODEL OPTION LOGGER RS 422 OPTION LOGGER access by the upper cover RS 422 WALL MOUNTING MODEL...

Page 5: ...perature S1 threshold NO contact Measurement Temperature S3 threshold NO contact Measurement Temperature NOT USED Ground N Power supply N Neutral L Live L NC mA BAMOX 7200 sensor mA 1 2 3 4 5 6 7 8 9...

Page 6: ...this key all parameters are saved previous data programmation will be overwrited If you are not sure of any modification do not press the VALID key To change parameters of data capture Num ric input...

Page 7: ...CATION MEASUREMENT PARAMETERS ATMOSPHERIC PRESSURE SALINITY ADJUST ALARM 1 ADJUST ALARM 2 ADJUST ALARM 3 RELAY REGULATION REGUL PID OUTPUT mA Oxygen OUTPUT mA TEMP TEMPERATURE FORCED RELAY ADJUST ALAR...

Page 8: ...tely the instrument acts within the configuration thresholds regulation analog outputs When PID regulation is activated the display shows the PID FORCED PID 00 00 mg l 0 000 one digit is flashing Modi...

Page 9: ...00 00 C Example Real value 22 C Reference 20 C Entry 2 00 C SAVING ATMOSPHERIC PRESSURE P 1013 hPa The blinding digit has to be modified if necessary Confirm with Enter then save SAVING SALINITY g kg...

Page 10: ...1 SAVING ADJUST ALARM 2 ADJUST ALARM 3 please see page 11 ALARM 2 ON OFF ALARM 2 MEASURE TEMP MEASURE Threshold against pH mV measured value TEMP Threshold against temperature measured value ALARM 2 L...

Page 11: ...ntact closes when the value goes under the limit ON 00 00 mg l C To close the contact S3 at this value OFF 00 00 mg l C To open the contact S3 at this value DELAY UP ONI OFF Delay or no delay before t...

Page 12: ...int T CYCLE 0000 SEC Cycle time Highest limit for HIGH PB 00 00 mg l proportional bandwith Lowest value LOW PB 00 00 mg l for dead zone Highest value H IGH DZ 00 00 mg l for dead zone Lowest value LOW...

Page 13: ...e corresponding timing Determinate on this graph both t ant T t delay of response T Time corresponding to the same variation in of measurement m and the analogic output Cde m Cde This value may be fou...

Page 14: ...mA TEMP TEMPERATURE HIGHER 0000 C Value corresponding to 20 00 mA LOWER 0000 C Value corresponding to 00 00 or 04 00 mA OUTPUT 4 20 mA 0 20 mA Output type SAVING L L L L L ENTER ENTER ENTER ENTER VALI...

Page 15: ...or too small When Error message occurs press MENU to be back to CONFIGURATION Proceed to a new calibration sequence If the ERROR message appears again cancel the calibration sequence go back to measu...

Page 16: ...LID MENU FORCED RELAY ADJUST ALARM ALARM 1 ON OFF ALARM 2 ON OFF Diagnostic mode to test the threshold configurations ALARM 3 ON OFF ALARM 4 ON OFF L L L L ENTER ENTER ENTER ENTER VALID MENU ADJUST AL...

Reviews: