Measured value
Measured value
Measured value
Measured value
Time
Time
Time
Time
RISING SIGNAL
DECREASING SIGNAL
Trigger point, relay ON
Trigger point, relay OFF
Trigger point, relay OFF
Trigger point, relay ON
Actuated relay
Actuated relay
Actuated relay
Actuated relay
Delay before relay switches ON
Delay before relay switches OFF
9.4
SETTING THE TEMPERATURE
Go to the menu TEMPERATURE
MEASURE : AUTO / MANUAL
AUTO : Temperature is displayed if a Pt 100 sensor is connected, confirm, then save.
MANUAL : Without Pt 100 sensor connected, enter the temperature manually.
Confirm
FLUID T. +000.0 °C
In the MANUAL mode, enter the temperature of the liquid, then confirm.
SAVE ?
To record the settings, press the icon SAVE
9.5
mA OUTPUT SETTINGS FOR THE MEASUREMENT
Go to the menu OUTPUT mA TURB
HIGHER 2000
Enter the measured value for which the output is 20 mA, then confirm.
LOWER 0000
Enter the measured value for which the output is 0 or 4 mA, then confirm
OUTPUT 4/20 mA or 0/20 mA
Select the output signal type, then confirm
SAVE ?
To record the settings, press the icon SAVE
9.6
mA OUTPUT SETTINGS FOR THE TEMPERATURE
Go to the menu OUTPUT mA TEMP.
160.0 °C
Enter the temperature value for which the output is 20 mA, then confirm.
LOWER +000.0 °C
Enter the temperature value for which the output is 0 or 4 mA, then confirm.
OUTPUT 4/20 mA or 0/20 mA
Select the output signal type, then confirm
SAVE ?
To record the settings, press the icon SAVE
9.7
SIMULATION OF RELAYS
With the menu FORCED RELAY, it is possible to manually test (force) relays S1, S2, S3 and S4.
Last one, S4, is N.O. by default. The test begins by relay S1.
The respective relay can be switched from OFF (open) to ON (closed).
Confirm to go to the next relay, then to go back to previous menu.
9.8
SIMULATION OF MEASUREMENT
Go to the menu FORCED MEASURE
0.000
Simulation at this measured value.
First displayed value is the current measurement.
Confirm, then enter the value you wish to test.
Note: The value is immediately considered by the monitor and actuates thresholds, 4-20mA outputs, etc.
TUR
07-06-2021
M-436.02-EN-AE
436-02
/11