Removing/ reinstalling the front screen
7.2
Invivo
2
Plus is fitted with a removable front screen, to facilitate easier cleaning and installation/
removal of equipment to and from the workstation chamber.
To remove the front screen;
•
Remove all cells/ samples to an alternative storage facility.
•
Switch the workstation off at the mains and remove the plug from the mains.
•
Remove the Ezee Sleeves
(see section 7.4.2 for more details).
•
Remove the front screen knobs by turning them anti-clockwise (counter-clockwise) to
remove them.
•
Disconnect the glove port selector knob using the quick release connector.
•
Remove the front screen by lifting it towards you. Place the front screen on a suitable
surface that will not scratch the front screen.
Take care not to lose any of the front cover knobs.
To reinstall the front screen;
•
Check that the front screen seals are clean. Clean the front screen seals using an approved
cleaning agent if required.
•
Place the front screen onto the mounting studs.
•
Reconnect the glove port selector knob using the quick release connector.
•
Tighten the front screen knobs onto the mounting studs by turning the front screen knobs
clockwise. Take care not to cross thread or over tighten the front screen knobs. The front
screen knobs should be tight enough to compress the seals.
•
Reinstall the Ezee Sleeves
(see section 7.4.2, for more details).
Cleaning procedure – deep clean
7.3
To deep clean the workstation;
Preparing the workstation
•
Remove all cells/ samples to an alternative storage facility.
•
Switch the workstation off at the mains and remove the plug from the mains.
•
Remove the front screen as described in section 7.2.
•
Remove any other equipment installed in the workstation chamber.
Cleaning the workstation chamber
•
The items inside the workstation chamber that require cleaning are;
•
The floor tray.
•
The workstation floor.
•
The ceiling panel.
•
The rear shelf.
•
The inner rear wall.