Page 32 of 61
InvivO
2
Initial Settings Screen
Pressing the “Atmosphere Control” bar on Screen 4 or Screen 5 will open up the settings
screen shown in Screen 6.
Screen 6: InvivO
2
Initial Settings Screen with Ultrasonic Humidity Option
Viewing the rows from left to right they display:
Row 1
Row 2
Row 3
Row 4
Measurements
Set-Points
High Alarm Levels
Low Alarm Levels
Table 6: Settings screen column functions
Rows 2, 3, and 4 can be changed by the user by pressing on the value and a pop-up will
appear for the new setting to be selected, as shown in Screen 7 for the Oxygen set-point
change
Screen 7: Oxygen Level Set-point Pop-up
Once all of the set-points are correct then pressing the arrow in the bottom right of the
screen will start the gas control.
Pressing the ‘Home’ button on the bottom left will return to the main menu.