398D003 Rev C
13 of 15
Step 5
: Remove the towel guard and remove power cord plug and M12 male signal connector.
Feed all cables from bottom of cabinet through the bottom plenum into the designated cable
glands. Afterwards re-attach M12 male signal connector and power cord plug.
Connect touch screen cables to appropriate locations on the rear of the InoqulA PC.
Step 6
: Install all removed cabinet parts in reverse order starting with blower, front seal panel,
viewscreen and weight box, touch screen mounts & printer cover and right side access panel
and outer side panel.
Step 7
: Replace 1/4-20 button head screws in view screen guides and front dress panels with
1/4-20 thumb screws
Towel guard
Cable port for touch screen cables
Power cable gland
Signal cable gland
1/4-20 button head
shipping screw
1/4-20 thumb screw