Safety equipment warnings and CE compliance statements
2/24/2010 | 71-018 V11 EN
D12R, D6R, D3R Users Manual
8
3)
Upon completion of a DC High Potential, Megohm, Polarization Index, Step Voltage,
Dielectric absorption, or Continuous Ramp test, before disconnecting the test leads,
short the winding, motor, etc., to ground and allow time for discharge. If this is not
done, voltage may still be active on leads and tested components.
4)
Make sure
the tester leads are disconnected before the motor is energized or
powered up.
5)
Do not
remove the product covers or panels or operate the tester without the
covers and panels properly installed. Components on inside of tester carry voltage
for operation and if touched can render a shock.
6)
Use appropriate safety equipment required by your organization, including high
voltage globes and eye protection.
7)
Repair Parts Warning
: Defective, damaged, or broken test leads must be
replaced with factory-authorized parts to ensure safe operation and maintain
performance specifications.
8)
Ground the product :
This product is grounded through the grounding conductor
of the power cord. To avoid electrical shock, plug the power cord into a properly
wired/grounded receptacle before connecting the product test leads.
Danger from loss of ground – Upon loss of the protective ground connection,
all accessible conductive parts, including knobs and controls that may appear
to be insulated, can cause an electric shock!
9)
This instrument is
NOT
waterproof or sealed against water entry.
10)
The unit is for indoor use. If used outdoors, the unit must be protected from rain,
snow and other contaminants.
Symbols on equipment
Protective conductor terminal. Located beside black ground test lead
on front panel of instrument.
Earth (ground) terminal
Frame or chassis terminal. Located on rear panel of instrument by
ground terminal.
Warning about hazardous voltage and risk of severe electrical shock or
death. Located beside each red test lead on front panel of instrument.
Other Information
Cleaning & decontamination
The D3R/D6R/D12R should be kept clean and in a dry environment. To clean the unit, power
down and unplug the instrument. Wipe with a clean water dampened cloth. Do not submerge