6.2 Operation and performance specifications
6.2.1 Flow rate ranges
Maximum and minimum flow rates through the
Sentinel LCT8
ultrasonic liquid flow transmitter are based on the pipe
diameter and the process fluid pressure. The information in
Table 18
below is approximate, and it is based on representative
liquid components at a process temperature of 70°F (21°C).
Table 18: Flow rate ranges
Nominal pipe size
Flow velocity
Volumetric flow
ft/s
m/s
m
3
/hr
gpm
bbl/hr
in.
mm
V
min
V
max
V
min
V
max
Q
min
Q
max
Q
min
Q
max
Q
min
Q
max
6
150
1
40
0.3
12.2
20
818
90
3602
129
5147
8
200
1
40
0.3
12.2
35
1417
156
6237
223
8913
10
250
1
40
0.3
12.2
56
2233
246
9831
351
14049
12
300
1
40
0.3
12.2
80
3203
353
14100
504
20150
14
350
1
40
0.3
12.2
98
3905
430
17191
614
24566
16
400
1
40
0.3
12.2
129
5172
569
22772
814
32542
18
450
1
40
0.3
12.2
165
6618
728
29137
1041
41637
20
500
1
40
0.3
12.2
206
8241
907
36285
1296
51852
24
600
1
40
0.3
12.2
301
12022
1323
52932
1891
75639
6.2.2 Operating pressures
The maximum operating pressures for
Sentinel LCT8
are listed in
Table 19
below.
Table 19: Maximum operating pressures by material and pressure class
Maximum operating pressures (psig)
Flange
class
WCB
LCB
CF8
CF8M
100°F
284°F
100°F
284°F
100°F
284°F
100°F
284°F
150#
285
235
265
234
275
209
275
218
300#
740
659
695
643
720
550
720
570
600#
1480
1316
1395
1282
1440
1095
1440
1139
Maximum operating pressures (barg)
Flange
class
WCB
LCB
CF8
CF8M
38°C
140°C
38°C
140°C
38°C
140°C
38°C
140°C
150#
19.6
16.2
18.3
16.1
19.0
14.4
19.0
15.0
300#
51.0
45.4
47.9
44.3
49.6
37.9
49.6
39.3
600#
102.0
90.7
96.1
88.4
99.2
75.5
99.2
78.5
NOTE:
For all load and force requirements and limitations, please consult Panametrics.