Chapter 2. Installation
18
Moisture Target Series 6 User’s Manual
2.6.2 Connecting a VeriDri Probe
Use the following steps to wire the
VeriDri
to the
MTS6
.
Figure 33: Four-Wire, VeriDri Probe Cable
1.
Insert the end of probe cable with the connector onto the probe and twist the connector head clockwise until it is
secure.
IMPORTANT:
Ensure that the power is off before proceeding.
2.
Using the flying leads at the end of the
VeriDri
cable, connect the probe to the lower terminal block (pins 10, 11, 12
and 13) on the back of the
MTS6
(see Table 1 and Figure 35 and Figure 36 on page 19).
IMPORTANT:
To maintain good contact at the terminal block and to avoid damaging the pins on the wiring
connector, pull the connector straight off (not at an angle) the terminal block. Then, make the cable
connections while the connector is off the unit. Finally, after the wiring is complete, push the connector
straight onto the terminal block (not at an angle).
Table 1: VeriDri Wiring Connections
Wire Color
Pin Number
Designation
Brown
10
RTN
Black
11
COM –
White
12
COM +
Blue
13
+15V
Summary of Contents for Panametrics MTS6
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 23: ...Moisture Target Series 6 User s Manual 15 Chapter 2 Installation Figure 27 MTS6 Wiring Diagram...
Page 32: ...Chapter 2 Installation 24 Moisture Target Series 6 User s Manual...
Page 73: ...Moisture Target Series 6 User s Manual 65 Chapter 4 Service and Maintenance...
Page 78: ...Chapter 5 Specifications 70 Moisture Target Series 6 User s Manual...
Page 83: ...Chapter B Menu Maps Moisture Target Series 6 Hygrometer User s Manual93 Chapter B Menu Maps...
Page 98: ...76 Moisture Target Series 6 User s Manual...
Page 104: ...Warranty 2 Moisture Target Series 6 User s Manual no content intended for this page...
Page 105: ......