background image

3.3  Cleaning the moisture probe

To clean your HygroPro moisture probe, carefully follow the instructions in this section.

3.3.1  Preparing to clean the probe

CAUTION!

Be sure to perform the probe cleaning 

procedure in a well-ventilated area. Observe 

all necessary safety precautions when 

handling the cleaning solvents

To clean the moisture probe, the following items are required:
•  Two glass (NOT metal) containers with approximately  

300 ml of reagent-grade hexane or toluene in each 

container.

•  One glass (NOT metal) container with approximately  

300 ml of distilled (NOT deionized) water.

IMPORTANT

Make sure the containers are deep enough to completely 

submerge the moisture probe. Do not place the 

transmitter module into the solvents. Insert only the 

moisture sensor into the solvents.

•  Rubber or latex gloves
•  An oven set at 50°C ± 2°C (122°F ± 3.6°F)
•  1-1/8” wrench

3.3.2  Replacing the RTE

To maximize HygroPro performance, Panametrics 

recommends recalibration of the aluminum oxide moisture 

sensor on the replaceable transducer element (RTE) every  

6 to 12 months. The optimum interval depends on the 

specific application. To accomplish this, either return  

the RTE to Panametrics for recalibration or install a new  

RTE. The HygroPro electronics will automatically read and  

store the calibration data whenever a new or recalibrated 

RTE is installed.

IMPORTANT

The programmed probe calibration data should not be 

changed without consulting Panametrics.

3.3.3  Removing the transmitter from the system

Complete the following steps to remove the transmitter from 

the installation site:
1.  Refer to "Figure 2" on page 3 and use a 1-1/8” wrench 

on the probe hex nut to unthread the transmitter from the 

fitting on the sample system or process line.

2.  Record the dew point of the ambient air.
3.  Disconnect the cable from the transmitter module.

3.3.4  Removing the probe from the transmitter

To remove the probe from the transmitter, refer to "Figure 9" 

on page 17 and proceed as follows:
1.  Loosen the four captive screws on the bottom of  

the transmitter.

2.  Carefully remove the metal plate without touching  

the sensor.

3.  Carefully pull the probe out of the transmitter.
4.  Disconnect the probe cable by turning the locknut at  

the top of the probe. Then, detach the sensor.

Captive screws

Cable locknut

Probe

Metal plate

Figure 9: Removing the probe from the transmitter

17

Summary of Contents for Panametrics HygroPro

Page 1: ...HygroPro User s manual 916 099 Rev D...

Page 2: ...ii...

Page 3: ...HygroPro Moisture transmitter User s manual 916 099 Rev D November 2012 iii...

Page 4: ...no content intended for this page proceed to next page iv...

Page 5: ...1 2 2 3 Selecting measurement parameters 11 2 2 4 Setting up the display 12 2 2 5 Setting up the analog output 12 2 2 6 Entering the node ID 13 2 2 7 Setting up the RS485 output 13 2 3 Advanced setup...

Page 6: ...the cleaned probe 18 Chapter 4 Specifications 19 4 1 General 19 4 2 Electrical 19 4 3 Mechanical 19 4 4 Moisture sensor 19 4 5 Built in temperature sensor 20 4 6 Built in pressure sensor 20 4 7 Certi...

Page 7: ...t the following link https www bakerhughesds com sales terms conditions Typographical conventions Note These paragraphs provide information that provides a deeper understanding of the situation but is...

Page 8: ...ke sure that unauthorized personnel cannot gain access to the operation of the equipment Environmental compliance RoHS The HygroPro moisture transmitter fully complies with RoHS regulations Directive...

Page 9: ...x button keypad and it is housed in an IP67 Type 4X enclosure The HygroPro transmitter uses an aluminum oxide moisture sensor and it includes a temperature thermistor and a pressure transducer on a co...

Page 10: ...he sample flow Before constructing a sample system consult a Panametrics application engineer and observe the following guidelines The sample system should be very simple and should contain as few com...

Page 11: ...age during operation 2 Using the integral 3 4 16 straight male thread screw the probe end of the transmitter into the process or sample system fitting Make sure not to damage the threads Note A 3 4 16...

Page 12: ...receive and display the 4 20 mA measurement output from the HygroPro proceed to next page an external power supply to provide power to the HygroPro plus a personal computer PC running Panametrics Pan...

Page 13: ...ground 1 Connect an RS485 RS232 converter customer supplied to an available serial port on the PC 2 Push the female connector end of the factory supplied cable into the mating male connector on the t...

Page 14: ...da for the following hazardous classified locations and protection methods Intrinsically safe class I II III division 1 group ABCDEFG T4 T Ambient 40 C to 60 C Enclosure IP 67 3 Galvanically isolated...

Page 15: ...C22 2 No 61010 1 Degrees of Protection Provided by Enclosures IP Code ANSI IEC 60529 Degrees of Protection Provided by Enclosures IP Code CSA C22 2 No 60529 1 4 2b Compliance requirements The HygroPr...

Page 16: ...e end of the HygroPro ATEX certificate number Bas06ATEX0019X indicates that special conditions are required for safe operation in Europe These conditions are 1 The equipment must be protected against...

Page 17: ...plied to the unit The transmitter requires up to 60 seconds to initialize and begin normal operation The unit will meet its specified accuracy within 3 minutes Figure 7 below shows a close up view of...

Page 18: ...st temperature Node ID enter value Baud rate Parity Stop bits Data Bits Save Table RS485 Service Zero mV Zero PSIg Span mV Span PSIg Temp C 6 CS 6 Temp C 1 CS 1 Row 16 Units 16 Row 1 Units 1 Figure 8...

Page 19: ...level passcode must be entered 2 2 3 Selecting measurement parameters To select the measurement parameters complete the following steps To select the number of measurements to be displayed on each scr...

Page 20: ...he above screen highlight that choice and press enter Use the arrow keys to change the contrast value and press enter Then press escape twice to return to the main menu 2 2 5 Setting up the analog out...

Page 21: ...twork when used with PanaView interface software To enter your nNode ID proceed as follows Enter your passcode to access the setup program Then use the arrow keys to select the Node ID menu Press ente...

Page 22: ...o enter the desired pressure value and press enter The screen will be updated to show the new pressure value To set the temperature display enter your password and use the arrow keys to select const t...

Page 23: ...to select lock keypad on the initial screen and press enter twice The keypad is now locked To unlock the keypad just re enter the user program as described on page 11 No further programming steps are...

Page 24: ...creen instructions After completing the above step respond to the ARE YOU SURE question Be sure to respond yes if you wish to save the new information 3 2 Moisture probe error conditions IMPORTANT All...

Page 25: ...al depends on the specific application To accomplish this either return the RTE to Panametrics for recalibration or install a new RTE The HygroPro electronics will automatically read and store the cal...

Page 26: ...n set at 50 C 2 C 122 F 3 6 F for 24 hours 3 3 6 Installing the probe in the transmitter To install a new or cleaned probe in the transmitter complete the following steps 1 Reconnect the probe cable t...

Page 27: ...iven a 24 VDC power supply Max load resistance 24 x 33 33 300 500 Cable 6 ft 2 m standard consult Panametrics for custom lengths Input parameters for loop powered intrinsic safety Ui 28V Pi 0 653W Li...

Page 28: ...lable ranges 30 to 300 psig 3 to 21 bar 50 to 500 psig 4 to 35 bar 100 to 1000 psig 7 to 69 bar 300 to 3000 psig 21 to 207 bar 500 to 5000 psig 35 to 345 bar Note Psig ranges are based on a constant p...

Page 29: ...s arising from course of dealing or usage or trade Return policy If a Panametrics instrument malfunctions within the warranty period the following procedure must be completed 1 Notify Panametrics givi...

Page 30: ...uction targets across critical industries including Oil Gas Energy Healthcare Water and Wastewater Chemical Processing Food Beverage and many others Join the conversation and follow us on LinkedIn lin...

Reviews: