Copyright 2011 Baker Hughes Company.
42 | DPS 8000 Instruction Manual–English
Appendix
A.
CANopen Object Dictionary
A.3.4 Object 6120h: AI Input Scaling 1 FV
This object defines the field value of the first calibration point for the analogue input channel. It is
scaled in physical unit of field value.
OBJECT DESCRIPTION
INDEX
6120
Name
AI Input scaling 1 FV
Object Code
Array
Data Type
REAL32
ENTRY DESCRIPTION
Sub-Index
000
Description
number of entries
Access
RO
PDO Mapping
No
Default Value
0.0
Sub-Index
001
Description
AI Input scaling 1 FV 1
Data Type
REAL32
Access
RW
PDO Mapping
No
Default Value
0.0
Sub-Index
002
Description
AI Input scaling 1 FV 2
Data Type
REAL32
Access
RW
PDO Mapping
No
Default Value
0.0
A.3.5 Object 6121h: AI Input Scaling 1 PV
This object defines the process value of the first calibration point for the analogue input channel.
It is scaled in physical unit of the process value.
OBJECT DESCRIPTION
INDEX
6121
Name
AI Input scaling 1 PV
Object Code
Array
Data Type
REAL32
ENTRY DESCRIPTION
Sub-Index
000
Description
number of entries
Access
RO
PDO Mapping
No
Default Value
0.0
Summary of Contents for Druck DPS 8000 Series
Page 1: ...Druck com DPS 8000 Series CAN bus Pressure Transducer Instruction Manual ...