4. Specification
4.1 Operating Environment
WARNING
This equipment is not rated for use in potentially explosive
atmospheres.
Parameter
Value
Storage temperatures
ADTS542F: -20°C to 70°C (-4° to 158°F)
ADTS552F )
ADTS553F ) -20°C to 70°C (-4° to 158°F)
ADTS554F )
Operating
temperatures/environment
ADTS542F: -5°C to 50°C (32° to 122°F)
ADTS552F )
ADTS553F ) -5°C to 50°C (32° to 122°F)
ADTS554F )
Operating humidity: 5% to 95% RH (condensing)
Indoor & Outdoor (non-continuously) use
ADTSTOUCH universal power
supply
Operating temperature: 10°C to 50°C (50° to 122°F)
Operating humidity: 5% to 95% RH (non-condensing)
Power supply
Universal input Mains Power Supply.
Class 1 equipment requiring a protective earth connection.
TXAH 250V HRC Time-delay (T) 250
Va.c. rated fuses (X = amp rating
of fuse).
Overvoltage category II, Pollution degree 2
The supplied mains/PSU cable protective earth/ground conductor must
be connected to the power supply protective earth/ground system.
Electrical safety
EN
61010-1, UL
61010-1, CSA 22.2, No. 61010-1 and IEC
61010-1
Operating altitude
Maximum 7500
ft (2300 metres)
Ingress protection
ADTS542F/552F/553F/554F IP33, ADTSTOUCH IP65 (EN60529)
Summary of Contents for Druck ADTS542F
Page 2: ......
Page 150: ...2014 Baker Hughes 148 ADTS542F 552F 553F 554F ADTSTOUCH 5 ADTSTOUCH...
Page 152: ...2014 Baker Hughes 150 ADTS542F 552F 553F 554F 7 4 4a 4a 4b 4c 4d...
Page 158: ...2014 Baker Hughes 156 ADTS542F 552F 553F 554F...
Page 165: ...2014 Baker Hughes Company ADTS542F 552F 553F 554F 163 ADTSTOUCH 5 ADTSTOUCH...
Page 172: ...2014 Baker Hughes Company 170 ADTS542F 552F 553F 554F...
Page 180: ...Copyright 2014 Baker Hughes Company 178 ADTS542F 552F 553F 554F ADTSTOUCH 5 ADTSTOUCH...
Page 182: ...Copyright 2014 Baker Hughes Company 180 ADTS542F 552F 553F 554F 7 4 4a 4a 4b 4c 4d...
Page 188: ...Copyright 2014 Baker Hughes Company 186 ADTS542F 552F 553F 554F...
Page 189: ......