SELECTING A MOUNTING LOCATION

Install COIN elements horizontally, rotated 90° along the pipe center line, see 

Figure 2

 . This method of mounting allows for free 

passage of solids and eliminates air entrapment at the transmitter connection . Other positions are acceptable provided there 

is proper venting of the transmitter and differences in lead line elevations are considered . Vertical installations as shown in 

Figure 3

 may introduce a slight hydrostatic head effect, which must be considered when zeroing the transmitter . Connections 

between the COIN meter and the transmitter should be 3/8 in . (9 .53 mm) tubing minimum . Use a three-valve manifold for 

zeroing the transmitter .

COINMETER

3-VALVE 

MANIFOLD

TRANSMITTER

CAUTION

Hd

Figure 2:   COIN meter with flanged taps

Figure 3:   Vertical installation

Each flow element has an arrow indicating the required direction of flow . Failure to properly orient the COIN element may 

cause improper results when using data supplied for an element that has been calibrated .

NOTES:

1. Φ =1 in./ft (80 mm/m) for water;

            2...4 in./ft (60...320 mm/m for more viscous fluids.

2. Minimize all lead line lengths.

Figure 4:  Typical horizontal installation for liquid

Selecting a Mounting Location  

Page 7 

March 2016

DPM-UM-00519-EN-04

Summary of Contents for PRESO COIN

Page 1: ...Flow Meters Model COIN User Manual DPM UM 00519 EN 04 March 2016...

Page 2: ...Flow Meters Model COIN Page ii March 2016 DPM UM 00519 EN 04...

Page 3: ...ormation 5 Unpacking and Inspection 5 Straight Pipe Run Requirements 6 Selecting a Mounting Location 7 Line Installation 10 Differential Pressure Connections 10 Typical Pressure Port Configurations 11...

Page 4: ...Flow Meters Model COIN Page iv March 2016 DPM UM 00519 EN 04...

Page 5: ...installation instructions for the Preso COIN flow meter to help you get the meter up and running quickly Safety Information The installation of the COIN flow meter must comply with all applicable fede...

Page 6: ...Figure 1 Piping requirements 1 For upstream and downstream lengths equal to one half the values shown above add 1 percent to the accuracy value 2 Any flow conditioner shall be installed in the straigh...

Page 7: ...ter Connections between the COIN meter and the transmitter should be 3 8 in 9 53 mm tubing minimum Use a three valve manifold for zeroing the transmitter COINMETER 3 VALVE MANIFOLD TRANSMITTER CAUTION...

Page 8: ...0 mm m for water 2 Minimize all lead line lengths Figure 5 Typical horizontal installation for gas Figure 6 Typical horizontal installation for steam Selecting a Mounting Location Page 8 March 2016 DP...

Page 9: ...Figure 7 Typical vertical installation for liquid Figure 8 Typical vertical installation for gas Selecting a Mounting Location Page 9 March 2016 DPM UM 00519 EN 04...

Page 10: ...EXCEEDING PROPER PRESSURE OR TEMPERATURE RATINGS CAN LEAD TO PERSONAL INJURY OR EQUIPMENT DAMAGE THE PROCESS PIPING FLANGES SHOULD BE IDENTICAL TO THOSE ON THE COIN METER THE PROCESS TEMPERATURE AND P...

Page 11: ...Do not exceed the specified torque DIFFERENTIAL PRESSURE CONNECTIONS The high pressure connection is always on the upstream side of the flow direction arrow and the Iow pressure connection on the down...

Page 12: ...ai UAE 971 4 371 2503 Czech Republic Badger Meter Czech Republic s r o Ma kova 2082 26 621 00 Brno Czech Republic 420 5 41420411 Slovakia Badger Meter Slovakia s r o Racianska 109 B 831 02 Bratislava...

Reviews: