background image

LM OG-AM series 

Oval gear meters with MID approved register and integrated 
datalogger 

LM_OG-AM_BA_02_1163 (May 2019)

 

User Manual 

LM OG-AM 

LM OG-TAERM 

LM OG-CNDAM 

Summary of Contents for Oval Gear LM OG-AM

Page 1: ...LM OG AM series Oval gear meters with MID approved register and integrated datalogger LM_OG AM_BA_02_1163 May 2019 User Manual LM OG AM LM OG TAERM LM OG CNDAM...

Page 2: ...conform meters 5 5 8 Installation procedure LM OG 5 6 Programming of LM OG AM and LM OG TAERM series 6 7 Calculating the correction factor 7 7 1 Calculation based default settings 7 7 2 Recalculation...

Page 3: ...table surface that may allow it to fall Never place the units above a radiator or heating unit Route all cabling away from potential hazards Isolate from the mains before removing any covers Power con...

Page 4: ...ation match with those of the lube meter for example connections pressure flow range and medium Once the meter has been installed please make sure that no air pressure shocks or particles can damage t...

Page 5: ...technicians or the officials of the national authorities MID 4 WARRANTY Badger Meter warrants meters and parts manufactured by it and supplied hereunder to be free from defects in materials and workma...

Page 6: ...front of the register No data are lost they are securely saved in an EEPROM memory except of the date and time To keep the date and time settings you have to reinsert the battery within 30 minutes See...

Page 7: ...dispense system or by an insufficiently areated oil pipe pressure shocks This can be prevented by installing an appropriate non return valve FLDIR will indicate a reverse flow quantity display will ch...

Page 8: ...arrow for the flow direction flashes now Use RESET to change or TOTAL to move forward 03 The measuring unit will flash and can be changed by pressing the RESET button L GAL QT PT To move forward pres...

Page 9: ...ctor would be for example 0 9960 Calculation 0 9960 x 1 003 0 9989 The new correction factor is 0 9989 8 PULSE OUTPUT AT REGISTER OG TAERM The pulse output on the LM OG TAERM register has a rating of...

Page 10: ...or more than 2s will bring you to the next lower menu level Pressing the TOTAL button once will bring you to the next newer data entry no matter if you are on the Date Time or Quantity menu level If y...

Page 11: ...e register will automatically move back to the standard display without saving the last entry 10 DISPLAY CORRECTION FACTOR To verify the programmed correction factor press TOTAL and RESET for 2 second...

Page 12: ...wake up the register Picture 4 Insert the battery cover then screw the battery cover tight Battery Lithium CR123A 3A33 07 L AC001 TOTAL 3A33 Checksum over the MID relevant source code 07 Sequential ve...

Page 13: ...ll CE declarations applying to this meter series under www badgermeter de en downloads ce declarations 16 DIN ISO AND EMC CERTIFICATE Please refer to our valid DIN ISO and EMC certificate under www ba...

Page 14: ...Return of goods for repair Harmlessness declaration Page 12 LM_OG AM_BA_02_1163 May 2019...

Page 15: ...Return of goods for repair Harmlessness declaration May 2019 LM_OG AM_BA_02_1163 Page 13...

Page 16: ...www badgermeter de...

Reviews: