Badger Meter ModMAG M2000 User Manual Download Page 13

Technical data 

Page 13/14 

 

 MID_M2000_BA_MB_02_1805 

  

5. 

Technical data 

The ModMAG

®

 M-Bus module is providing a EN13757 compatible M-Bus interface to 

the Badger ModMAG

® 

M2000 flow

 

meter 

 

 

Product 

ModMAG

®

 M2000 M-Bus module 

Power supply 

2 x +5V from M2000 internal module interface  

(+5 V digital, +5V field) 

Current 

1 mA from +5V digital 

10 mA frp, +5V field 

ModMAG

®

 M2000 

interface 

2 wire ModBus

®

 interface 

9600, 19200, 38400 baud auto-baud detection at firmware 

start 

8 data bits,  

1 stop bit, 

1 even parity bit 

Phoenix IMC 1.5/11-G-3.81 connector to M2000 

M-Bus interface 

2 wire EN13757 compatible M-Bus interface 

300, 2400, 9600 baud auto-baud detection 

8 data bits 

1 stop bit 

1 even parity bit 

1 M-Bus unit load (1.5 mA) 

15 mA active M-Bus current 

M-Bus input with reversible mains protection and protective 

earth clamp 

3 pin clamp 

Isolation 

2500 V RMS isolation between M-Bus interface and 

ModMAG

®

 M2000 

Module size 

54 x 37 mm, maximum height 27 mm 

Operating temperature  -20 °C … +60 °C (-4 °F … +140 °F) 
Storage temperature 

-40 °C … +70 °C (-40 °F … +160 °F) 

Humidity 

90 % 

Product life time 

˃ 10 years 

 
 
 
 
 
 

Summary of Contents for ModMAG M2000

Page 1: ...Badger Meter Europa GmbH ModMAG M2000 M Bus interface User manual May 2018 MID_M2000_BA_MB_02_1805...

Page 2: ...ary address 6 4 5 Changing baud rate 6 4 6 Changing M Bus response telegram 7 4 7 Write configuration area to flash 7 4 8 Send ModBus commands 8 4 9 M Bus REQ_UD2 answers 8 4 9 1 M Bus REQ_UD2 answer...

Page 3: ...primary address is saved in a non volatile memory 300 2400 and 9600 baud communication speed Automatic baud rate detection or manually programmable baud rate Standard M Bus serial communication parame...

Page 4: ...in board The interface board is supported by a pad to the wall of the enclosure A grounding strap connect the interface board terminal 63 to the nearest screw of the main board 3 1 Electrical connecti...

Page 5: ...e addressed using its primary address range 0 250 The default factory setting primary address of the module is 0 zero The primary address can be reconfigured using the appropriate M Bus command see be...

Page 6: ...urs time 4 5 Changing baud rate By default factory setting the module is automatically set to detect the baud rate of the incoming M Bus request telegram Usually it is not necessary to change this set...

Page 7: ...s are then answered with the selected telegram Please note that the RSP_UD telegram setting is not written immediately in the non volatile flash memory of the module but only On the cyclic 24 hours re...

Page 8: ...Signature 0F ModBus ChkS 16 e g 68 1D 1D 68 08 00 72 95 09 10 19 A9 09 01 07 08 01 00 00 Header for M Bus RSP_UD 0F Flag manufacturer specific 01 03 0A 31 39 31 30 30 39 39 35 00 00 ModBus answer Chk...

Page 9: ...for handling readout requests see also EN13757 3 chapter 8 3 4 9 1 M Bus REQ_UD2 answer All N ModBus register Description 0 0x00CF T1 T in m3 1 0x00D7 T2 T in m3 2 0x00D7 T3 TN in m3 3 0x00E7 T1 T Ro...

Page 10: ...03 Inst Volume m3 1 1 0 0 REAL4 0 000000e 000 Inst Volume m3 2 2 0 0 REAL4 1 854350e 003 Inst Volume m3 3 0 0 0 REAL4 0 000000e 000 Inst Volume Flow 1 sec 10E3 4 0 0 0 INT2 0 Inst Control signal N Uni...

Page 11: ...er factor 7 0x007B FACTORY Amplifier factor 8 0x007D Detector current in mA 9 0x007F FACTORY Detector current in mA 10 0x0081 Power line frequency in Hz 11 0x0082 Excitation frequency in Hz 12 0x010B...

Page 12: ...major version 12 0x0059 Daugtherboard minor version 13 0x011C Security status N Unit Tariff Storage Data Value Funct VIB 0 0 0 0 INT2 1 Inst Model Version 1 0 0 0 Var M2000 Inst Model Version 2 0 0 1...

Page 13: ...ction at firmware start 8 data bits 1 stop bit 1 even parity bit Phoenix IMC 1 5 11 G 3 81 connector to M2000 M Bus interface 2 wire EN13757 compatible M Bus interface 300 2400 9600 baud auto baud det...

Page 14: ...Harmlessness declaration Page 14 14 MID_M2000_BA_MB_02_1805 6 Return of goods for repair Harmlessness declaration Please refer to our claims return form harmlessness declaration under www badgermeter...

Page 15: ......

Page 16: ...N rtinger Strasse 76 72639 Neuffen Germany E mail badger badgermeter de www badgermeter de Durchflu richtung r ckw Durchflu richtung r ckw Stromausg ang aktiv 0 4 20mA Vorw rts Normal offen Normal ges...

Reviews: