Parameters
Page 33
August 2019
DPP-UM-00256-EN-03
PAOAMETEOS
The following tables list and explain the parameters The tables are divided into three main sections:
• Setup parameters
• Module configuration parameters
• Advanced setup parameters
Setup Parameters
The setup menu contains all of the configuration parameters necessary to make the transmitter operational
Parameter
Function
Options
Description
ID UNITS
Measurement
standard
ENGLSH (Inches)
METRIC (Millimeters)
Selects the unit of measure for pipe ID entry
ID
Pipe inside
diameter
(Enter a numeric value) Enter the pipe inside diameter in inches if
ENGLSH
was selected as
UNITS
; in millimeters if
METRIC
was selected
DISPLAY
Display mode
RATE
TOTAL
BOTH
DIA
To display only the Flow Rate, select
OATE
To display only the Flow Total, select
TOTAL
To
alternately display the Flow Rate and the Total, select
BOTH
By selecting BOTH, the display
switches between RATE and TOTAL every 7 seconds
The
DIA
selection places the display in the diagnostics mode When selected, the display shows
the measured frequency, the gain setting and the signal strength
RATE UNT
Engineering units
for flow rate
Select an engineering unit for flow rate measurements When Pounds (LBS) or Kilograms (KGS) is selected, the specific
gravity for the fluid type must be entered for the
SP GOAV
setup parameter
Measurement
Selection
Measurement
Selection
Pounds
LBS
Cubic Meters
CUBIC ME
Kilograms
KGS
Millions of Metric Liters
MEGLTRS
Gallons
GALLONS
Acre Feet
ACRE FT
Liters
LITERS
Oil Barrels (42 Gallons)
OIL BARR
Millions of Gallons
MGAL
Liquor Barrels (31 5 Gallons)
LIQ BARR
Cubic Feet
CUBIC FT
Velocity in Linear Feet
VEL FEET
Millions of Cubic Feet
M CU FT
Velocity in Linear Meters
VEL MTRS
RATE INT
Time interval for
flow rate
SEC Seconds
MIN Minutes
HOUR Hours
DAY Days
Select a time interval for flow rate measurements
TOTL UNT
Totalizer units
Select an engineering unit for flow totalizer measurements When Pounds (LBS) or Kilograms (KGS) is selected, the specific
gravity for the fluid type must be entered for the
SP GOAV
setup parameter
Measurement
Selection
Measurement
Selection
Pounds
LBS
Cubic Meters
CUBIC ME
Kilograms
KGS
Millions of Metric Liters
MEGLTRS
Gallons
GALLONS
Acre Feet
ACRE FT
Liters
LITERS
Oil Barrels (42 Gallons)
OIL BARR
Millions of Gallons
MGAL
Liquor Barrels (31 5 Gallons)
LIQ BARR
Cubic Feet
CUBIC FT
Velocity in Linear Feet
VEL FEET
Millions of Cubic Feet
M CU FT
Velocity in Linear Meters
VEL MTRS
TOTL MUL
Flow totalizer
multiplier value
0 01…1,000,000
Used for setting the flow totalizer exponent This feature is useful for accommodating a very
large accumulated flow The exponent is a ×10
n
multiplier, where “n” can be from -2 (×0 01) to +6
(×1,000,000) Reference the table below for valid entries and their influence on the DFX meter
display
Exponent
Display Multiplier
× PT 01 (-2)
× 0 01
× PT 1 (-1)
× 0 1
×1 (0)
× 1
×10 (1)
× 10
×100 (2)
× 100
×1000 (3)
× 1000
×10000 (4)
× 10,000
×100000 (5)
× 100,000
×1000000 (6)
× 1,000,000
SP GRAV
Fluid specific
gravity
(Enter a numeric value)
Allows adjustments to be made to the specific gravity (density relative to water) of the liquid
If Pounds (LBS) or Kilograms (KGS) is selected for either the RATE UNT or the TOTL UNT, a specific
gravity must be entered for the correct mass flow to be calculated A list of fluids and their
associated specific gravities is located in
FL C-OFF
Low flow cutoff
(Enter a numeric value)
A Low Flow Cutoff entry is provided to allow very low flow rates (that can be present when
pumps are off and valves are closed) to be displayed as Zero flow The value entered is in
actual
rate units