Installation
Page 8
June 2020
TRB-UM-00287-EN-07
5 Pipe Diameters
Minimum
10 Pipe Diameters
Minimum
Isolation
Valve
QuikSert Turbine
Flow Meter
Figure 4: Meter installation without using a bypass line
This is true with any restriction in the flow line that may cause the liquid to flash�
If necessary, air eliminators should be installed to ensure that the meter is not
incorrectly measuring entrained air or gas�
It is recommended that a minimum length, equal to ten (10) pipe diameters of
straight pipe, be installed on the upstream side and five (5) diameters on the
downstream side of the flow meter� Otherwise, meter accuracy may be affected�
Piping should be the same size as the meter bore or threaded port size�
Do not locate the flow meter or connection cable close to electric motors,
transformers, sparking devices, high voltage lines, or place connecting cable in
conduit with wires furnishing power for such devices� These devices can induce
false signals in the flow meter coil or cable, causing the meter to read inaccurately�
If problems arise with the flow meter and monitor, consult
(Pickup Not Included)" on page 13
� If further problems arise, consult the factory�
If the internal components of the turbine flow meter are damaged beyond repair,
turbine meter repair kits are available� Information pertaining to the turbine meter
repair kits is referenced in
"Repair Kit Information" on page 14
�