Service Kit Parts Installation
Page 4
April 2021
DSY-UM-02594-EN-04
Electrical Symbols
Function
Direct
Current
Alternating
Current
Earth
(Ground)
Protective
Ground
Chassis
Ground
Symbol
SERVICE KIT PARTS INSTALLATION
DANGER
REMOVE (DE-ENERGIZE) ANY EXTERNAL POWER FROM THE UNIT BEFORE REMOVING
THE SCREW COVER FROM THE ENCLOSURE. FAILURE TO DO SO CAN BE DANGEROUS.
Service Kit Number B290-901 contains:
•
One programmed PCB assembly
•
Four screws, Phillips pan head, #4 × 1/2, zinc/steel
1� Loosen the 2 enclosure captive screws and open the cover�
2� Remove the four screws holding the board in place�
3� Tilt the board slightly to access the ribbon cable�
4� Remove the ribbon cable from the J4 connector on the display side of the board�
5� Press the tab on the battery connector plug to release it from the P2 battery lead
connector socket�
6� Cut and discard the tie wrap that holds the battery�
OTE:
N
You will need to replace the tie wrap that holds the battery in place�
7� Remove the battery and set it aside�