Badger Meter 735 Series User Manual Download Page 4

INTOODUCTION

Used in conjunction with any Badger Meter® flow monitor or endpoint, Badger Meter non-magnetic Impeller flow sensors 

provide an accurate reading of the rate of liquid flow as well as total accumulated flow  A number of sensor models are 

offered, which cover applications for a wide range of pipe sizes and pressure/temperature specifications 
The flow sensors generate a frequency which is proportional to flow rate  An internal preamplifier allows the pulse signal to 

travel up to 2000 feet without further amplification  The impeller bearing assembly, shaft and O-rings are replaceable in  

the field  
The Series 735 Impeller flow sensors feature a four-blade impeller design, using a proprietary, non-magnetic sensing technology  

As the liquid flow turns the impeller, a low impedance signal is transmitted with a frequency proportional to the flow rate  
Sensors of similar type are interchangeable, so there is no need for recalibration after servicing or replacement  

ELECTOONIC TYPES

Sensors are supplied with two single-conductor 18 AWG solid copper wire leads with UL Style 116666 direct burial insulation  

These IR models are used in below grade applications such as irrigation, municipal and groundwater monitoring  All 

Badger

 

Meter Series 735 sensor electrical components are self-contained  See 

"Specifications" on page 10

 for pressure and 

temperature ratings 

IO Sensor

Designed for below grade applications such as irrigation, municipal, and groundwater monitoring where the flow rates are 

between 2…20 ft/sec  and temperatures are below 110° F  IR sensors are supplied with two single conductor, 18 AWG solid 

copper wire with UL Style 116666 direct burial insulation 

Series 735

These models feature a modified PVC tee with solvent weld socket end connections, and a removable PPS sensor insert  Available 

sizes are 1/2 in  (12 7 mm), 3/4 in  (19 05 mm) and 1 in  (25 4 mm)  BSP threaded adapters are available 

MECHANICAL INSTALLATION 

DEPRESSURIZE AND VENT PIPING SYSTEM PRIOR TO ANY INSTALLATION OR MAINTENANCE OF THE FLOW SENSOR.

The accuracy of flow measurement for all flow measuring devices is highly dependent on proper location of the sensor in the 

piping system  Irregular flow velocity profiles caused by valves, fittings, and pipe bends can lead to inaccurate overall flow 

rate indications, even though local flow velocity measurement may be accurate  A sensor located in the pipe where it can be 

affected by air bubbles, floating debris or sediment may not achieve full accuracy and could be damaged  Badger Meter flow 

sensors are designed to operate reliably under adverse conditions, but the following recommendations should be followed to 

ensure maximum system accuracy:
•  Choose a location along the pipe with ten pipe diameters upstream and five pipe diameters downstream of the sensor 

provide no flow disturbance  Pipe bends, valves, other fittings, pipe enlargements and reductions should not be present in 

this length of pipe 

•  The preferred location for the sensor around the circumference of a horizontal pipe is on top  If trapped air or debris will 

interfere, then the sensor should be located further around the pipe from the top but not more than 45 degrees from 

top dead center  The sensor should never be located at the bottom of the pipe, as sediment may collect there  Locations 

off top dead center cause the impeller friction to increase, which may affect performance at low flow rates and increase 

wear  Any circumferential location is correct for installation in vertical pipes  Rising flow preferred to reduce effects of any 

trapped air 

Introduction

Page 4 

October 2015

SEN-UM-01665-EN-04

Summary of Contents for 735 Series

Page 1: ...User Manual Sensor Series 735 Plastic Tee Type Flow Sensor SEN UM 01665 EN 04 October 2015 ...

Page 2: ...lectronic Types 4 IR Sensor 4 Series 735 4 Mechanical Installation 4 Electrical Installation Guidelines 5 Flow Sensor Installation 6 Additional Information 7 Calibration 8 Calibration Table 8 Dimensions 9 Specifications 10 Ordering Matrix 11 Sensor Series 735 Plastic Tee Type Flow Sensor Page ii October 2015 SEN UM 01665 EN 04 ...

Page 3: ...uld result in severe personal injury or death Indicates a hazardous situation which if not avoided is estimated to be capable of causing minor or moderate personal injury or damage to property Considerations THE IMPELLER STYLE FLOW SENSOR DESCRIBED IN THIS MANUAL IS NOT INTENDED FOR USE IN SAFETY CRITICAL APPLICATIONS USE OF THE DEVICE IN THIS MANNER IS DONE AT THE SOLE DISCRETION OF THE CUSTOMER ...

Page 4: ...eature a modified PVC tee with solvent weld socket end connections and a removable PPS sensor insert Available sizes are 1 2 in 12 7 mm 3 4 in 19 05 mm and 1 in 25 4 mm BSP threaded adapters are available MECHANICAL INSTALLATION DEPRESSURIZE AND VENT PIPING SYSTEM PRIOR TO ANY INSTALLATION OR MAINTENANCE OF THE FLOW SENSOR The accuracy of flow measurement for all flow measuring devices is highly d...

Page 5: ...ke sure the epoxy is hardened before inverting the splice or dropping it in standing water Do not make an underground splice unless absolutely necessary Route the cable from the sensor to a Badger Meter flow monitor or endpoint The cable may be extended up to 2000 ft using two conductor shielded 20 AWG or larger stranded copper wire with appropriate ratings Leave enough flexibility in the cable or...

Page 6: ...ate primer and cement according to the manufacturer s instructions Do not use or leave excess glue Glue build up in the tee can affect the measurement performance of the unit and or prevent the impeller from turning Do not let the glue or primer contact the bore where the O rings seal 4 Insert the sensor into the tee carefully lining up the locating hole in the sensor with the locating pin in the ...

Page 7: ...edule 40 Keep all parts as clean as possible Do not let the glue or primer contact the surfaces where the O rings make their seal Although the meter can measure at velocities up to 20 ft sec do not exceed velocities above 7 5 ft sec Flow Sensor Installation Page 7 October 2015 SEN UM 01665 EN 04 ...

Page 8: ...ibration Table Size Schedule K Offset 1 2 S40 0 078000 0 9 1 2 SDR 13 5 0 120119 0 1 3 4 S40 0 156300 0 9 3 4 SDR 21 0 197000 0 6 1 S40 0 261119 1 2 1 SDR 21 0 321739 0 6 Input the information in the table above into the following equation to determine the frequency of the Series 735 Impeller flow sensors Freq gpm K offset This information is required when calibrating an output board when using th...

Page 9: ...P Optional 1 2 in Ø 0 840 3 85 in 97 8 mm 3 06 in 77 7 mm 6 086 in 154 6 mm 3 4 in Ø 1 050 3 85 in 97 8 mm 3 31 in 84 1 mm 6 775 in 172 1 mm 1 in Ø 1 315 3 94 in 100 1 mm 3 50 in 88 9 mm 6 775 in 172 1 mm Dimensions Page 9 October 2015 SEN UM 01665 EN 04 ...

Page 10: ...ll scale over recommended design flow range Repeatability 1 5 of full scale over recommended design flow range Linearity 1 5 of full scale over recommended design flow range Transducer Excitation 8 35V DC max input source limited to 100 mA Quiescent current 600 uA 8 35V DC max Quiescent voltage Vhigh Supply Voltage 600 uA Supply impedance ON State Vlow Max 1 2V DC 40 mA current limit 15 Ω 0 7V DC ...

Page 11: ... 1 Sch 40 only PV SIZE 0 5 05 0 75 07 1 10 0 5 with BSP Adapters 06 0 75 with BSP Adapters 08 1 with BSP Adapters 11 ELECTRONICS HOUSING PPS 0 ELECTRONICS IR Irrigation 6 O RING EPDM 1 SHAFT Tungsten Carbide 2 IMPELLER 300 SST 0 BEARING UHMWPE 1 Ordering Matrix Page 11 October 2015 SEN UM 01665 EN 04 ...

Page 12: ...eter de las Americas S A de C V Pedro Luis Ogazón N 32 Esq Angelina N 24 Colonia Guadalupe Inn CP 01050 México DF México 52 55 5662 0882 Europe Middle East and Africa Badger Meter Europa GmbH Nurtinger Str 76 72639 Neuffen Germany 49 7025 9208 0 Europe Middle East Branch Office Badger Meter Europe PO Box 341442 Dubai Silicon Oasis Head Quarter Building Wing C Office C209 Dubai UAE 971 4 371 2503 C...

Reviews: