R•22 id
Instruction 2100-9002
6
2.
Turn on the power switch located on the top of the unit. For battery operation,
press the switch to “BAT.” For auxiliary power operation, connect the optional
auxiliary power supply to the bottom of the unit and press the switch to “AUX.”
3.
The lights will sequence, press “NEXT” to begin the “WARM UP.”
4.
After approximately 90 seconds, the “CALIBRATING” light will fl ash.
5. Press
“NEXT.”
6.
The “CALIBRATING” light will illuminate for approximately 60 seconds.
7.
When the “ANALYZING” indicator begins to fl ash, connect hose to the system or
cylinder’s low side service port and then press “NEXT.”
8.
The “ANALYZING” indicator will illuminate for approximately 45 seconds while the
test is in progress.
9.
The “PASS”/”FAIL” light will then provide the test results (See Below)
10. Disconnect the sample hose from the test sample, press the “NEXT” button and
allow the unit to complete a calibration cycle before powering off the unit. The
instrument may now be stowed.
2.3 Interpreting the Results
After the Analysis is complete, the “PASS” or “FAIL” light will fl ash.
1.
“PASS” indicates the refrigerant tested is 95% or greater R22.
2.
“FAIL” indicates the refrigerant tested is less than 95% R22 and should not be
recovered without special equipment.
3.
The EXCESS AIR” light will illuminate in conjunction with the “PASS” or “FAIL” if
the instrument determines that a signifi cant amount of air is present. (See Trouble-
shooting) .