Portable Gas Monitor (PGM-IR) Infrared Sensor Replacement
4
Instruction 3015-5508 (Rev 2)
Figure 5. Monitor Components in Various PGM-IR Models
Step
Description
8.
Disconnect the tubing from the purge-air bag “T”-fitting, and then lay aside the
chassis-half containing the purge-air bag. (This step does not apply to models without
the purge air bladder bag.)
9.
Pull the 10-pin and 12-pin electrical connectors from their sockets on the IR sensor.
Note that these connectors may have been secured in place for shipping purposes
with hot-glue or wire ties. As necessary, carefully pry the hot-glue from each
connector or cut off the wire ties.
10.
Pull the tubing from the gas-inlet and gas-outlet fittings on the IR sensor.
11.
Using wire cutters, cut the two cable ties that secure the IR sensor to the chassis, and
then remove the sensor.