Bacharach PGM-IR Operation And Maintenance Download Page 12

Operation 

CO

2

 Monitor 

Instruction 3015-9000 Rev 2

 

2.2 

Connecting Gas

-

Sample

 & Exhaust Lines 

2.2.1 

Connecting the Gas

-

Sample 

Probe 

To connect the gas-sample probe to the monitor, simply push the elbow connector into the Gas Sample Port on 

the front of the monitor as shown in the illustration below.  

 

The gas-sample line can be up to 

50 ft (15.2 m)

 in length. All tubing bends should have a radius of no less than 5” 

to ensure proper airflow. If kinks or obstructions occur in the line the monitor may not function properly.

  

When searching for leaks, the end of the probe should be placed near the potential leak source and positioned to 

reduce the possibility of mists, aerosols, oil, water, dust, or other contaminates being drawn into the monitor.  

2.2.2 

Connecting the Exhaust Line

 (Optional) 

The exhaust line carries the exhausted gas sample away from the monitor, and should be installed if the monitor is 

located in a confined, poorly ventilated area.  
The exhaust line can be up to 

50 ft (15.2 m)

 in length. Ideally this line should terminate outdoors in a location that 

is not exposed to the elements.  

2.2.3 

Using the Push

-to-

Connect (PTC) Fittings

 

Connect the exhaust line to the monitor by firmly pushing the tubing into the Exhaust Port’s push-to-connect (PTC) 

fitting.  See below.  If the exhaust line terminates outside the building, position the tubing so that no water or 

moisture can enter the line.  Note that this line does not require a termination filter.   

 

To connect the exhaust line to the exhaust 

port’s PTC  fitting, firmly push the 

appropriate tubing into the hole in the 

center of the connector until it seats in 

the connector.  Refer to the figures at the 

left. 
To remove tubing from a PTC connector, 

push and hold the spring-loaded collar 

inwards, then simultaneously withdraw 

the tubing.  Refer to the figures at the 

right. 

 

 

Summary of Contents for PGM-IR

Page 1: ...PGM IR CO2 Monitor Portable Gas Monitor for Carbon Dioxide CO2 Gas Instruction 3015 9000 Operation and Maintenance Rev 2 April 2014 Patent 6 590 690 Product Leadership Training Service Reliability...

Page 2: ...CO2 Monitor ii Instruction 3015 9000 Rev 2...

Page 3: ...ARE GIVEN AND ACCEPTED IN LIEU OF I ANY AND ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND II...

Page 4: ...CO2 Monitor iv Instruction 3015 9000 Rev 2...

Page 5: ...Peak Reading Operation 8 2 6 Leak Detector Operation 8 2 7 Standby Mode 8 2 8 Battery Charging and Using 9 2 8 1 Charging the Battery 9 2 8 2 Using the Battery 9 2 9 Display Screens 10 2 9 1 Initial...

Page 6: ...Monitor vi Instruction 3015 9000 Rev 2 3 2 Internal Particulate Hydrophobic Filter Replacement 22 3 3 Updating Firmware 22 4 PARTS AND SERVICE 23 4 1 Replacement Parts 23 4 2 Optional Accessories 23 4...

Page 7: ...d associated with the use of this equipment It calls attention to a procedure practice or condition or the like which if not correctly performed or adhered to could result in personal injury or death...

Page 8: ...r may be impaired if the monitor is used in a manner not specified by these instructions Changes or modifications to this monitor will void the warranty IMPORTANT This device is designed for operation...

Page 9: ...n audible alarm are provided to signal gas alarm and instrument fault conditions The audible alarm can temporarily be silenced by a push of a button but will reactivate if the gas level is still above...

Page 10: ...idity 5 to 90 RH non condensing Altitude Limit 6 562 ft 2 000 m Power DC power pack provides a minimum of 8 hours of operation Power Consumption 15 Watts maximum Front panel 3 indicator lights ON Gree...

Page 11: ...CO2 Monitor Operation Instruction 3015 9000 Rev 2 5 2 Operation 2 1 Front Panel Display Controls...

Page 12: ...the exhausted gas sample away from the monitor and should be installed if the monitor is located in a confined poorly ventilated area The exhaust line can be up to 50 ft 15 2 m in length Ideally this...

Page 13: ...SILENCE function page 13 If a system fault occurs see Fault Code list on page 14 the monitor responds by turning ON the front panel FAULT yellow light If the AUDALRM function is activated page 11 the...

Page 14: ...ng the 2ndUNIT function The display above shows that currently the monitor is detecting 50 ppm CO2 and that a peak measurement of 70 ppm has occurred The peak reading is reset by pressing the ENTER bu...

Page 15: ...ed However you should fully charge the battery prior to operating the PGM IR Refer to the battery manufacturer s instruction sheet for the proper charging procedure 2 8 2 Using the Battery To properly...

Page 16: ...is at the same temperature and humidity as the area being checked for gas DO NOT fill the bag in an area that is contaminated with gas The following message is displayed and an audible alarm is sound...

Page 17: ...ON and the audible alarm to sound if AUDALRM function is set to ALARM or ALARM FAULT AUDALRM The internal audible alarm is programmable to function as follows OFF Audible alarm is disabled ALARM Alar...

Page 18: ...desired time period and then press ENTER to save that value and return to the previous screen P CHK The Pressure Check function displays additional diagnostic information that can be helpful when trou...

Page 19: ...continue to flash however as a reminder that a fault condition still exists The audible alarm circuit will reactivate at the end of the silence period if the cause of the fault has not been corrected...

Page 20: ...inate the purge air bag 0200 Gain Set Fault The digipot autotune sequence has failed Call the factory for further instructions 0400 A D Fault A fault has occurred in the analog to digital circuitry Co...

Page 21: ...ure This is the current internal enclosure temperature in C Fault Code Current fault code Pressure Reading This is the pressure as measured every purge cycle with the sample pump off and the gas sampl...

Page 22: ...rent manifold pressure in psia Stored Ambient Pressure Stored ambient pressure in psia Pressure Difference The difference between the current manifold pressure and the stored ambient pressure Fault Co...

Page 23: ...CO2 Monitor Operation Instruction 3015 9000 Rev 2 17...

Page 24: ...as follows a Click Start click Run In the Run box type hypertrm exe and click OK The New Connection dialog box should appear b Type in a New Connection Name Click OK to select the default icon if desi...

Page 25: ...re capable of importing comma delimited files The following describes how to generate a spreadsheet from a comma delimited text file using Microsoft Excel If you are using a different spreadsheet prog...

Page 26: ...Operation CO2 Monitor 20 Instruction 3015 9000 Rev 2 NOTE A worksheet must be open in order to import data A sample spreadsheet is shown below...

Page 27: ...or s metal chassis as follows Items Required Medium Phillips head screwdriver Procedure 1 Unplug battery pack from monitor 2 Remove the monitor from the soft carrying case Remove a total of 10 screws...

Page 28: ...illustration below and locate the internal particulate hydrophobic filter to be replaced 3 Pull off tubing from both ends of filter and remove filter from instrument 4 Attach tubing to new filter 5 R...

Page 29: ...art Number IrDA Printer includes printer roll of thermal paper 4 x AA batteries manual 0024 1400 Thermal printer paper 1 roll 0006 8733 Thermal printer paper 5 rolls 0024 1310 Cable DB9 to RJ 45 3015...

Page 30: ...3015 9000 Rev 2 Headquarters 621 Hunt Valley Circle New Kensington PA 15068 PH 724 334 5000 Toll Free 800 736 4666 FAX 724 334 5001 Website www mybacharach com E mail help mybacharach com Printed in...

Reviews: